Impact of inter-crystal scattering on image quality of a dedicated brain PET scanner with DOI capability

<i>Objective</i>. Smaller scintillator crystals are commonly used in positron emission tomography (PET) to achieve higher resolution for preclinical research and clinical diagnosis, particularly in brain-dedicated PET scanners with depth-of-interaction (DOI) capability. Although DOI information provides parallax error correction, mispositioning of the line-of-response (LOR) induced by inter-crystal scattering (ICS) can lead to performance degradation, affecting both spatial resolution and image contrast. Previous studies primarily focused on ICS elimination or recovery on PET scanners that are equipped with one-to-one coupled scintillation detectors, with few studies proposing applicable methods for handling ICS events in brain-dedicated PET scanner equipped with light-sharing window detectors and DOI capability. This study specifically evaluates the impact of ICS on a system that have already undergone DOI correction.<i>Approach</i>. By positioning ICS events to their first interaction crystal in Monte Carlo simulations and removing ICS events through a contour-based mask in experimental studies, evaluations of the impact of ICS events on image quality in the context of DOI-corrected systems were conducted. Metrics including spatial resolution and peak-to-valley ratio (PVR) were used for quantitative analysis, while reconstructed images and line profiles provided qualitative comparisons.<i>Main results</i>. Three stages of images were compared in both simulation and experimental studies, revealing that although image quality could be improved by DOI correction, the presence of ICS still affects spatial resolution and image contrast. Applying a contour-based ICS removal method prior to image reconstruction showed a higher PVR, resulting in improved spatial resolution and better lesion detectability for the scanner, though at the cost of sensitivity loss.<i>Significance</i>. Through simulation and experimental evaluation, we have demonstrated that the presence of ICS can severely degrade the performance of high-resolution brain-dedicated PET scanners with DOI decoding capabilities, highlighting the necessity for accurate ICS event classification.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-11-05
DOI
https://doi.org/10.1088/1361-6560/ae1bf2
Primary Topic
Radiation Detection and Scintillator Technologies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Impact of inter-crystal scattering on image quality of a dedicated brain PET scanner with DOI capability

Jiajia Tao, Qiyu Peng, Haoyu Zou, Xin Yu et al.
Physics in Medicine and Biology
Radiation Detection and Scintillator Technologies
article

Impact of inter-crystal scattering on image quality of a dedicated brain PET scanner with DOI capability

Jiajia Tao, Qiyu Peng, Haoyu Zou, Xin Yu, Zhijun Zhao, Shuang Xie, Xiaoying Tang, Huiping Zhao, Da Liang, Yibin Zhang
article en

Abstract

<i>Objective</i>. Smaller scintillator crystals are commonly used in positron emission tomography (PET) to achieve higher resolution for preclinical research and clinical diagnosis, particularly in brain-dedicated PET scanners with depth-of-interaction (DOI) capability. Although DOI information provides parallax error correction, mispositioning of the line-of-response (LOR) induced by inter-crystal scattering (ICS) can lead to performance degradation, affecting both spatial resolution and image contrast. Previous studies primarily focused on ICS elimination or recovery on PET scanners that are equipped with one-to-one coupled scintillation detectors, with few studies proposing applicable methods for handling ICS events in brain-dedicated PET scanner equipped with light-sharing window detectors and DOI capability. This study specifically evaluates the impact of ICS on a system that have already undergone DOI correction.<i>Approach</i>. By positioning ICS events to their first interaction crystal in Monte Carlo simulations and removing ICS events through a contour-based mask in experimental studies, evaluations of the impact of ICS events on image quality in the context of DOI-corrected systems were conducted. Metrics including spatial resolution and peak-to-valley ratio (PVR) were used for quantitative analysis, while reconstructed images and line profiles provided qualitative comparisons.<i>Main results</i>. Three stages of images were compared in both simulation and experimental studies, revealing that although image quality could be improved by DOI correction, the presence of ICS still affects spatial resolution and image contrast. Applying a contour-based ICS removal method prior to image reconstruction showed a higher PVR, resulting in improved spatial resolution and better lesion detectability for the scanner, though at the cost of sensitivity loss.<i>Significance</i>. Through simulation and experimental evaluation, we have demonstrated that the presence of ICS can severely degrade the performance of high-resolution brain-dedicated PET scanners with DOI decoding capabilities, highlighting the necessity for accurate ICS event classification.

Physics in Medicine and BiologyVol. 70(23)
Southern University of Science and Technology (CN), Shenzhen Bay Laboratory (CN)
Openalex Percentile: Top 57%
Radiation Detection and Scintillator Technologies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.