There is an urgent need for validated tools to measure sclerotic cutaneous chronic graft-versus-host disease (scGVHD). We examined the interobserver reproducibility within a session and intraobserver repeatability between sessions of the Myoton device for quantifying skin sclerosis in 36 adults with scGVHD. The Myoton was used to measure oscillation frequency and relaxation time of soft tissues at 7 bilateral sites (14 anatomic sites) by 2 study personnel at 2 study sessions. Agreement was measured using mean pairwise absolute difference (MPD), and reliability was measured using intraclass correlation coefficient (ICC). For each of the 2 Myoton parameters, the overall interobserver MPD was <5% of the average overall values and the interobserver ICC was >0.90 between the 2 observers, indicating excellent agreement and reliability within a measurement session. The median time between sessions 1 and 2 was 47.5 days. The overall normalized intraobserver MPD was <7% of the average overall values for each of the 2 Myoton parameters, reflecting good agreement between sessions. The intraobserver ICC for frequency and relaxation time parameters were 0.85 and 0.84, respectively, indicating good reliability between sessions. The reproducibility and repeatability of a bonus site selected at each study visit were similar to the standard 14 anatomic sites. However, no individual site was nearly as reproducible or repeatable as the overall Myoton measurements averaged across the patient. Our findings emphasize the utility of the Myoton for assessing skin properties in scGVHD with patient-level measurements.

Authors

Institutions

Publication Details

Journal
Blood Advances
Published
2025-11-05
DOI
https://doi.org/10.1182/bloodadvances.2025016648
Citations
1
Primary Topic
Hematopoietic Stem Cell Transplantation
Type
article
Field-Weighted Citation Impact
1.77

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reproducibility and repeatability of the Myoton to quantify sclerotic chronic graft-versus-host disease

Najla El Jurdi, Jerry Mi, Nosha Farhadfar, Eric R. Tkaczyk et al.
1 citations
Blood Advances
Hematopoietic Stem Cell Transplantation
1.77
article

Reproducibility and repeatability of the Myoton to quantify sclerotic chronic graft-versus-host disease

Najla El Jurdi, Jerry Mi, Nosha Farhadfar, Eric R. Tkaczyk, Stephanie J. Lee, Mongoljin Bat-Erdene, Lacey Williams, Shramana Ghosh, Heidi Chen, Kelsey Baker, R. B. Sahu, Rachel B. Weiss, Gabriela Desatnik
article en
1 citations

Abstract

There is an urgent need for validated tools to measure sclerotic cutaneous chronic graft-versus-host disease (scGVHD). We examined the interobserver reproducibility within a session and intraobserver repeatability between sessions of the Myoton device for quantifying skin sclerosis in 36 adults with scGVHD. The Myoton was used to measure oscillation frequency and relaxation time of soft tissues at 7 bilateral sites (14 anatomic sites) by 2 study personnel at 2 study sessions. Agreement was measured using mean pairwise absolute difference (MPD), and reliability was measured using intraclass correlation coefficient (ICC). For each of the 2 Myoton parameters, the overall interobserver MPD was <5% of the average overall values and the interobserver ICC was >0.90 between the 2 observers, indicating excellent agreement and reliability within a measurement session. The median time between sessions 1 and 2 was 47.5 days. The overall normalized intraobserver MPD was <7% of the average overall values for each of the 2 Myoton parameters, reflecting good agreement between sessions. The intraobserver ICC for frequency and relaxation time parameters were 0.85 and 0.84, respectively, indicating good reliability between sessions. The reproducibility and repeatability of a bonus site selected at each study visit were similar to the standard 14 anatomic sites. However, no individual site was nearly as reproducible or repeatable as the overall Myoton measurements averaged across the patient. Our findings emphasize the utility of the Myoton for assessing skin properties in scGVHD with patient-level measurements.

Blood AdvancesVol. 10(4)
National Institutes of Health (US), Vanderbilt University (US), Knoxville College (US), Fred Hutch Cancer Center (US), Sarah Cannon (US), Methodist Hospital (US), Imaging Technology Consulting (United States) (US), National Cancer Institute (US), Department of Veterans Affairs (AU), Vanderbilt University Medical Center (US)
U.S. Department of Veterans Affairs, Pfizer, AstraZeneca, Sanofi, Incyte, Syndax Pharmaceuticals, National Institutes of Health
Openalex Percentile: Top 11%
Hematopoietic Stem Cell Transplantation
1.77
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.