Feasibility of radiation dose reduction with AI-based spectral reconstruction for computed tomography

Abstract Purpose The purpose of this study is to investigate whether 50% radiation dose with AI-based spectral reconstruction has equivalent or non-inferior image quality to routine dose with iterative reconstruction. Methods Fifteen patients were scanned at two radiation dose levels during a standard renal ablation procedure. Ablation planning images were obtained at routine dose (CTDI vol = 18.9 ± 4.7 mGy), and probe placement images were obtained at low dose (CTDI vol = 9.9 ± 2.5 mGy). Quantitative assessment included noise measurements and line profiles of iterative reconstruction (iDose, Philips) and AI-based spectral reconstruction (Spectral Precise Image, Philips) with respect to filtered-back-projection. Qualitative assessment included a radiologist reader study comparing routine dose images with iterative reconstruction (RD + IR) and low dose images with AI-based spectral reconstruction (LD + AI). Two radiologists rated overall image quality on a five-point Likert scale ranging from very poor (1) to excellent (5). A Wilcoxon signed-rank test was used to test for a significant difference between the reconstruction conditions. Results For conventional results, LD + AI was rated significantly higher than RD + IR ( p = 0.0051). For 50 keV MonoE and electron density results, there was not a significant difference between the LD and RD protocols. Of the results tested, the image quality of 50 keV was rated higher than conventional and electron density results. Conclusion This study suggests the feasibility of reducing radiation dose with AI-based spectral reconstruction without compromising overall image quality requirements for CT-guided renal ablations. For the conventional result, readers indicated superior image quality in LD + AI relative to RD + IR result.

Authors

Institutions

Publication Details

Journal
CVIR Oncology
Published
2025-11-05
DOI
https://doi.org/10.1007/s44343-025-00021-3
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Feasibility of radiation dose reduction with AI-based spectral reconstruction for computed tomography

Rebecca Hibbert, Nathan R. Huber, Ahmad Parvinian, Brian T. Welch et al.
CVIR Oncology
Radiation Dose and Imaging
article

Feasibility of radiation dose reduction with AI-based spectral reconstruction for computed tomography

Rebecca Hibbert, Nathan R. Huber, Ahmad Parvinian, Brian T. Welch, Christopher P. Favazza
article en

Abstract

Abstract Purpose The purpose of this study is to investigate whether 50% radiation dose with AI-based spectral reconstruction has equivalent or non-inferior image quality to routine dose with iterative reconstruction. Methods Fifteen patients were scanned at two radiation dose levels during a standard renal ablation procedure. Ablation planning images were obtained at routine dose (CTDI vol = 18.9 ± 4.7 mGy), and probe placement images were obtained at low dose (CTDI vol = 9.9 ± 2.5 mGy). Quantitative assessment included noise measurements and line profiles of iterative reconstruction (iDose, Philips) and AI-based spectral reconstruction (Spectral Precise Image, Philips) with respect to filtered-back-projection. Qualitative assessment included a radiologist reader study comparing routine dose images with iterative reconstruction (RD + IR) and low dose images with AI-based spectral reconstruction (LD + AI). Two radiologists rated overall image quality on a five-point Likert scale ranging from very poor (1) to excellent (5). A Wilcoxon signed-rank test was used to test for a significant difference between the reconstruction conditions. Results For conventional results, LD + AI was rated significantly higher than RD + IR ( p = 0.0051). For 50 keV MonoE and electron density results, there was not a significant difference between the LD and RD protocols. Of the results tested, the image quality of 50 keV was rated higher than conventional and electron density results. Conclusion This study suggests the feasibility of reducing radiation dose with AI-based spectral reconstruction without compromising overall image quality requirements for CT-guided renal ablations. For the conventional result, readers indicated superior image quality in LD + AI relative to RD + IR result.

CVIR OncologyVol. 1(1)
Philips (Finland) (FI), Mayo Clinic in Arizona (US)
Openalex Percentile: Top 53%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.