Carotid DECTA in patients with dental hardware: Impact of deep learning image reconstruction and metal artifact reduction

To evaluate the effectiveness of deep learning image reconstruction (DLIR) with metal artifact reduction (MAR) (DLIR-MAR) for improving carotid dual-energy CT angiography (DECTA) in patients with dental hardware. This retrospective study included 49 patients with dental hardware who underwent carotid DECTA. Virtual monochromatic images (VMIs) were reconstructed with DLIR-MAR, adaptive statistical iterative reconstruction-Veo (ASIR-V), and ASIR-V with MAR (ASIRV-MAR) at 40-keV and 50-keV energy levels. We quantitatively compared image noise, CT attenuation, contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and artifact index (AI) among different reconstructions, and performed a qualitative assessment of internal carotid artery (ICA) visualization affected by metal artifacts and overall image quality using a 5-point scale. Statistical analyses used repeated measures ANOVA or Friedman tests, with paired t-tests or Wilcoxon signed-rank tests as appropriate. Both ASIRV-MAR and DLIR-MAR reduced metal artifacts compared with ASIR-V, with lower AI and higher vascular visualization scores ( P < 0.001). Compared to ASIRV-MAR, DLIR-MAR reduced image noise at both keV levels ( P < 0.001). DLIR-MAR 40-keV VMIs exhibited comparable image noise ( P > 0.05), higher CT attenuation, SNR, CNR, and overall image quality compared with ASIRV-MAR 50-keV VMIs ( P < 0.05). Although DLIR-MAR 40-keV VMIs had higher AI values than ASIRV-MAR 50-keV VMIs ( P < 0.001), this did not significantly affect ICA visualization ( P > 0.05). DLIR reconstructed with MAR improves carotid DECTA images in patients with dental hardware, reducing image noise in lower-keV VMIs to capture their contrast enhancement benefits. • Metal artifact reduction (MAR) significantly reduces the metal artifacts in carotid dual-energy CT angiography (DECTA) images of patients with dental hardware. • Deep learning image reconstruction (DLIR) with MAR further reduces image noise in 40-keV virtual monochromatic images (VMIs) compared to iterative reconstruction with MAR. • DLIR with MAR ensures the use of 40-keV VMIs to further improve contrast enhancement for carotid DECTA.

Authors

Institutions

Publication Details

Journal
Meta-Radiology
Published
2025-11-05
DOI
https://doi.org/10.1016/j.metrad.2025.100188
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Carotid DECTA in patients with dental hardware: Impact of deep learning image reconstruction and metal artifact reduction

Z. Li, Rong‐Rong He, Yewen Jiang, Liucheng Li et al.
Meta-Radiology
Advanced X-ray and CT Imaging
article

Carotid DECTA in patients with dental hardware: Impact of deep learning image reconstruction and metal artifact reduction

Z. Li, Rong‐Rong He, Yewen Jiang, Liucheng Li, Yuguo Li, Z. M. Zheng, Shuangshuang Xu, Yu Yongqiang, Xiaohu Li, Yong Cheng, Zimeng Yeyan
article en

Abstract

To evaluate the effectiveness of deep learning image reconstruction (DLIR) with metal artifact reduction (MAR) (DLIR-MAR) for improving carotid dual-energy CT angiography (DECTA) in patients with dental hardware. This retrospective study included 49 patients with dental hardware who underwent carotid DECTA. Virtual monochromatic images (VMIs) were reconstructed with DLIR-MAR, adaptive statistical iterative reconstruction-Veo (ASIR-V), and ASIR-V with MAR (ASIRV-MAR) at 40-keV and 50-keV energy levels. We quantitatively compared image noise, CT attenuation, contrast-to-noise ratio (CNR), signal-to-noise ratio (SNR), and artifact index (AI) among different reconstructions, and performed a qualitative assessment of internal carotid artery (ICA) visualization affected by metal artifacts and overall image quality using a 5-point scale. Statistical analyses used repeated measures ANOVA or Friedman tests, with paired t-tests or Wilcoxon signed-rank tests as appropriate. Both ASIRV-MAR and DLIR-MAR reduced metal artifacts compared with ASIR-V, with lower AI and higher vascular visualization scores ( P < 0.001). Compared to ASIRV-MAR, DLIR-MAR reduced image noise at both keV levels ( P < 0.001). DLIR-MAR 40-keV VMIs exhibited comparable image noise ( P > 0.05), higher CT attenuation, SNR, CNR, and overall image quality compared with ASIRV-MAR 50-keV VMIs ( P < 0.05). Although DLIR-MAR 40-keV VMIs had higher AI values than ASIRV-MAR 50-keV VMIs ( P < 0.001), this did not significantly affect ICA visualization ( P > 0.05). DLIR reconstructed with MAR improves carotid DECTA images in patients with dental hardware, reducing image noise in lower-keV VMIs to capture their contrast enhancement benefits. • Metal artifact reduction (MAR) significantly reduces the metal artifacts in carotid dual-energy CT angiography (DECTA) images of patients with dental hardware. • Deep learning image reconstruction (DLIR) with MAR further reduces image noise in 40-keV virtual monochromatic images (VMIs) compared to iterative reconstruction with MAR. • DLIR with MAR ensures the use of 40-keV VMIs to further improve contrast enhancement for carotid DECTA.

Meta-RadiologyVol. 4(1)
Anhui Medical University (CN), First Affiliated Hospital of Anhui Medical University (CN)
Openalex Percentile: Top 68%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.