Validation of the CONTEC08C oscillometric blood pressure monitor in adults according to the International Organization for Standardization 81060-2:2018 standard

This study aimed to assess the accuracy of the CONTEC08C automated oscillometric upper-arm sphygmomanometer in adults for clinical use in accordance with the ISO 81060-2:2018 standard. A total of 90 normotensive and hypertensive participants aged 18-75 years were recruited, and 85 were analyzed to evaluate the accuracy of the CONTEC08C oscillometric upper-arm sphygmomanometer using the same arm sequential validation method and a standardized protocol. The mean ± standard deviation (SD) of the differences between the test device and reference blood pressure (BP) measurements was 0.5 ± 3 mmHg for SBP and 0.4 ± 3.5 mmHg for DBP, meeting validation criterion 1. For validation criterion 2, the SD of the averaged differences between the test device and reference BP measurements per subject was 1.81/2.41 mmHg (systolic/diastolic). The CONTEC08C oscillometric upper-arm blood pressure monitor fulfilled the requirements of the ISO 81060-2:2018 standard in adults in the clinical setting and can therefore be recommended for safe clinical use.

Authors

Institutions

Publication Details

Journal
Blood Pressure Monitoring
Published
2025-11-03
DOI
https://doi.org/10.1097/mbp.0000000000000781
Primary Topic
Blood Pressure and Hypertension Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of the CONTEC08C oscillometric blood pressure monitor in adults according to the International Organization for Standardization 81060-2:2018 standard

Wenjie Wang, Jin‐Song Bian, Jingjing Bai, Ying Lou et al.
Blood Pressure Monitoring
Blood Pressure and Hypertension Studies
article

Validation of the CONTEC08C oscillometric blood pressure monitor in adults according to the International Organization for Standardization 81060-2:2018 standard

Wenjie Wang, Jin‐Song Bian, Jingjing Bai, Ying Lou, Mengjia Chen, Lu Wang, Jun Cai, Wenjun Ma
article en

Abstract

This study aimed to assess the accuracy of the CONTEC08C automated oscillometric upper-arm sphygmomanometer in adults for clinical use in accordance with the ISO 81060-2:2018 standard. A total of 90 normotensive and hypertensive participants aged 18-75 years were recruited, and 85 were analyzed to evaluate the accuracy of the CONTEC08C oscillometric upper-arm sphygmomanometer using the same arm sequential validation method and a standardized protocol. The mean ± standard deviation (SD) of the differences between the test device and reference blood pressure (BP) measurements was 0.5 ± 3 mmHg for SBP and 0.4 ± 3.5 mmHg for DBP, meeting validation criterion 1. For validation criterion 2, the SD of the averaged differences between the test device and reference BP measurements per subject was 1.81/2.41 mmHg (systolic/diastolic). The CONTEC08C oscillometric upper-arm blood pressure monitor fulfilled the requirements of the ISO 81060-2:2018 standard in adults in the clinical setting and can therefore be recommended for safe clinical use.

Blood Pressure MonitoringVol. 31(2)
Chinese Academy of Medical Sciences & Peking Union Medical College (CN), Beijing Anzhen Hospital (CN), Fu Wai Hospital (CN)
Openalex Percentile: Top 50%
Blood Pressure and Hypertension Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.