Real-Time Volumetric Alignment for Image-Guided Brain Tumor Resection: A Dynamic Computational Framework
Intra-operative brain shift degrades the spatial accuracy of neuronavigation systems that rely exclusively on pre-operative magnetic resonance imaging (MRI). This study presents and evaluates a Dynamic Data-Driven Non-Rigid Registration (DDD-NRR) framework that couples multi-scale multimodal feature extraction, linear elastic biomechanical regularization, and temporal smoothness within a hybrid CPU–GPU pipeline to align pre-operative MRI with intra-operative 3D ultrasound (iUS). Across a 25-volume biomechanically simulated synthetic dataset (256×256×256 voxels), DDD-NRR achieved a mean Target Registration Error (TRE) of 1.23±0.41 mm, a Dice Similarity Coefficient (DSC) of 0.89±0.05, and a Hausdorff distance of 3.12±0.87 mm in 47.3±8.6 s per registration cycle, outperforming Biomechanical Finite Element Modeling (2.15±0.73 mm), GPU Demons (1.87±0.62 mm), and VoxelMorph (2.42±0.91 mm; repeated-measures ANOVA F(3,72)=41.6, p<0.001). In a retrospective clinical evaluation of 18 brain tumor resection cases (5 validated against intra-operative MRI and 13 evaluated via multi-expert anatomical landmark correspondence), the framework achieved a mean TRE of 1.64±0.19 mm (95% CI: [1.55, 1.74] mm) and a mean DSC of 0.88±0.02 (95% CI: [0.86, 0.89]). While performance remains constrained in cases with extensive resection cavities and severe acoustic shadowing, the framework demonstrates sub-2 mm technical registration accuracy within intra-operative time constraints.
Authors
- Latha Kiran Krishna Rajendran
Publication Details
- Journal
- International Journal of Clinical Research and Medical Sciences
- Published
- 2026-11-23
- DOI
- https://doi.org/10.67231/cyk4j317
- Citations
- 3
- Primary Topic
- Medical Image Segmentation Techniques
- Type
- article
- Field-Weighted Citation Impact
- 2.09