This work investigates the impact of scatter correction on photon-counting dual-energy chest radiography using a moving block method, focusing on quantifying improvements with the IEC 62220-2-1 dual-energy metrics. A modified LucAl-based chest phantom with PMMA and aluminum inserts was modeled in three sizes (small, standard, large) to represent different patient sizes. Monte Carlo simulations with MC-GPU and the Photon Counting Toolkit were used to simulate a CdTe photon-counting detector with two energy thresholds at 30 and 70 keV. Scatter was estimated from blocker shadows at 25 positions, interpolated across the field of view, and smoothed with a Gaussian filter (σ=5.0 mm), then subtracted separately from low- and high-energy images. Performance was evaluated using the per-feature dual-energy contrast (DEC) and the kerma-normalized dual-energy subtraction efficiency (DSE) with all acquisitions normalized to an entrance air kerma of 1 mGy to reflect clinical exposure conditions. In simulations, the moving block estimate reproduced the true scatter distribution with an average pixel-wise error of 0.4%. Scatter contamination introduced visible artifacts in the dual-energy subtraction images, particularly in aluminum-enhanced (Al-enhanced) images, and reduced contrast for target materials by up to 25%, as reflected in both DEC and DSE values at a fixed dose. Scatter correction restored image contrast, increased DEC for target materials while keeping non-target DEC low, and reduced edge artifacts across phantom sizes with the largest gains in the large phantom. These results support the moving block method as a dose-neutral strategy to improve dual-energy subtraction performance in photon-counting chest radiography.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-11-03
DOI
https://doi.org/10.3390/s25216734
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Scatter Removal in Photon-Counting Dual-Energy Chest X-Ray Imaging Using a Moving Block Method: A Simulation Phantom Study

Bahaa Ghammraoui, Yee Lam Elim Thompson
Sensors
Advanced X-ray and CT Imaging
article

Scatter Removal in Photon-Counting Dual-Energy Chest X-Ray Imaging Using a Moving Block Method: A Simulation Phantom Study

Bahaa Ghammraoui, Yee Lam Elim Thompson
article en

Abstract

This work investigates the impact of scatter correction on photon-counting dual-energy chest radiography using a moving block method, focusing on quantifying improvements with the IEC 62220-2-1 dual-energy metrics. A modified LucAl-based chest phantom with PMMA and aluminum inserts was modeled in three sizes (small, standard, large) to represent different patient sizes. Monte Carlo simulations with MC-GPU and the Photon Counting Toolkit were used to simulate a CdTe photon-counting detector with two energy thresholds at 30 and 70 keV. Scatter was estimated from blocker shadows at 25 positions, interpolated across the field of view, and smoothed with a Gaussian filter (σ=5.0 mm), then subtracted separately from low- and high-energy images. Performance was evaluated using the per-feature dual-energy contrast (DEC) and the kerma-normalized dual-energy subtraction efficiency (DSE) with all acquisitions normalized to an entrance air kerma of 1 mGy to reflect clinical exposure conditions. In simulations, the moving block estimate reproduced the true scatter distribution with an average pixel-wise error of 0.4%. Scatter contamination introduced visible artifacts in the dual-energy subtraction images, particularly in aluminum-enhanced (Al-enhanced) images, and reduced contrast for target materials by up to 25%, as reflected in both DEC and DSE values at a fixed dose. Scatter correction restored image contrast, increased DEC for target materials while keeping non-target DEC low, and reduced edge artifacts across phantom sizes with the largest gains in the large phantom. These results support the moving block method as a dose-neutral strategy to improve dual-energy subtraction performance in photon-counting chest radiography.

SensorsVol. 25(21)
United States Food and Drug Administration (US), Center for Devices and Radiological Health (US), Office of Science (US)
Openalex Percentile: Top 69%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Scatter Removal in Photon-Counting Dual-Energy Chest X-Ray Imaging Using a Moving Block Method: A Simulation Phantom Study — Bahaa Ghammraoui, Yee Lam Elim Thompson · Sensors (2025) | TGRS Research Map | TGRS