Comparison of Frictional Forces Among Ureteral Access Sheaths and Urinary Tract Dilators in a Novel <i>Ex Vivo</i> Porcine Ureteral Model

<b><i>Introduction:</i></b> Surface lubricity is a desirable characteristic of ureteral access sheaths (UAS), allowing for atraumatic navigation of the urinary tract. This study aimed to characterize the surface lubricity of commonly available 14F UAS, as well as 14F ureteral and urethral dilators, using a tribometer and a novel <i>ex vivo</i> ureteral tissue holder. <b><i>Materials and Methods:</i></b> The surface lubricity of 12, 14F urological catheters (8 UAS, 1 urethral dilator, and 3 ureteral dilators) was tested in six <i>ex vivo</i> Yorkshire porcine ureteral segments. A tribometer, capable of measuring force in thousandths of a Newton (N), measured frictional force during insertion and retraction cycles under a 4.0 N normal force. The coefficient of friction (COF) was calculated. Individual and cumulative lubricity scores (ILS and CLS) were calculated based on COF percentiles, enabling a comparative evaluation of device performance. Surface roughness was assessed via white light interferometry. <b><i>Results:</i></b> The Cook Flexor® UAS consistently demonstrated the highest lubricity, as reflected in its CLS for both insertion and retraction phases, followed closely by the Wellead ClearPetra® UAS. Dilators, including the Cook AQ hydrophilic-coated ureteral and urethral dilators, exhibited higher COF values, correlating with greater surface roughness. COF values were significantly higher during retraction cycles than insertion cycles (<i>p</i> < 0.05). Variability in COF across devices underscored the influence of surface properties on lubricity. <b><i>Conclusions:</i></b> Among all the UAS/dilators tested, the Cook Flexor® AQ-coated UAS and Wellead ClearPetra UAS had the best lubricity. Catheter retraction resulted in higher COF than catheter advancement.

Authors

Institutions

Publication Details

Journal
Journal of Endourology
Published
2025-11-03
DOI
https://doi.org/10.1177/08927790251389962
Primary Topic
Kidney Stones and Urolithiasis Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Comparison of Frictional Forces Among Ureteral Access Sheaths and Urinary Tract Dilators in a Novel Ex Vivo Porcine Ureteral Model

Antônio Rebello Horta Görgen, Pengbo Jiang, Alberto Martini, Ralph V. Clayman et al.
Journal of Endourology
Kidney Stones and Urolithiasis Treatments
article

Comparison of Frictional Forces Among Ureteral Access Sheaths and Urinary Tract Dilators in a Novel Ex Vivo Porcine Ureteral Model

Antônio Rebello Horta Görgen, Pengbo Jiang, Alberto Martini, Ralph V. Clayman, Bruce M. Gao, Roshan M. Patel, Jaime Altamirano, Ahmed S. Ali, Zachary E. Tano, Andrei D. Cumpanas, Thao N. Vu, Jaime Landman, Brandon Camp, Kevin Moreno-Ruiz
article en

Abstract

<b><i>Introduction:</i></b> Surface lubricity is a desirable characteristic of ureteral access sheaths (UAS), allowing for atraumatic navigation of the urinary tract. This study aimed to characterize the surface lubricity of commonly available 14F UAS, as well as 14F ureteral and urethral dilators, using a tribometer and a novel <i>ex vivo</i> ureteral tissue holder. <b><i>Materials and Methods:</i></b> The surface lubricity of 12, 14F urological catheters (8 UAS, 1 urethral dilator, and 3 ureteral dilators) was tested in six <i>ex vivo</i> Yorkshire porcine ureteral segments. A tribometer, capable of measuring force in thousandths of a Newton (N), measured frictional force during insertion and retraction cycles under a 4.0 N normal force. The coefficient of friction (COF) was calculated. Individual and cumulative lubricity scores (ILS and CLS) were calculated based on COF percentiles, enabling a comparative evaluation of device performance. Surface roughness was assessed via white light interferometry. <b><i>Results:</i></b> The Cook Flexor® UAS consistently demonstrated the highest lubricity, as reflected in its CLS for both insertion and retraction phases, followed closely by the Wellead ClearPetra® UAS. Dilators, including the Cook AQ hydrophilic-coated ureteral and urethral dilators, exhibited higher COF values, correlating with greater surface roughness. COF values were significantly higher during retraction cycles than insertion cycles (<i>p</i> < 0.05). Variability in COF across devices underscored the influence of surface properties on lubricity. <b><i>Conclusions:</i></b> Among all the UAS/dilators tested, the Cook Flexor® AQ-coated UAS and Wellead ClearPetra UAS had the best lubricity. Catheter retraction resulted in higher COF than catheter advancement.

Journal of EndourologyVol. 39(12)
University of California, Merced (US), University of California, Irvine (US)
Openalex Percentile: Top 54%
Kidney Stones and Urolithiasis Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.