Photon-Counting Computed Tomography of the Paranasal Sinuses Improves Intraoperative Accuracy of Image-Guided Surgery

<b>Background:</b> Computed tomography (CT)-based image-guided surgery (IGS) is of great importance in functional endoscopic sinus surgery (FESS) and requires IGS-specific imaging protocols to ensure high intraoperative accuracy. This study aimed to compare photon-counting CT (PCCT), dual-energy dual-source CT (DECT), and spectral detector CT (SDCT) of the paranasal sinuses with respect to image quality, IGS accuracy and radiation dose. <b>Methods:</b> A formalin-fixed cadaver skull was examined using PCCT, DECT and SDCT at 100 kV tube voltage with descending tube currents (mAs). The setup of electromagnetic IGS was evaluated using a visual analog scale. Accuracy was analyzed endoscopically using defined anatomical landmarks. Diagnostic image quality as well as bone and soft tissue noise were assessed qualitatively using a 5-point Likert scale and quantitatively by determination of signal-to-noise ratio. Radiation dose was evaluated using the dose length product. <b>Results:</b> While PCCT datasets could be registered and navigated accurately down to 10 mAs (1.5 mm error at 10 mAs), both DECT and SDCT exhibited significantly increased inaccuracies below 40 mAs (4.35/5.15 mm for DECT/SDCT at 25 mAs). Using PCCT therefore enabled a 45% radiation dose reduction at the minimally required dose length product using PCCT. Quantitative and qualitative image quality were superior for PCCT compared to DECT and SDCT. <b>Conclusions:</b> PCCT provides excellent accuracy of anatomical landmarks in IGS with superior image quality of the paranasal sinuses in low-mA scans and substantially reduced radiation exposure.

Authors

Institutions

Publication Details

Journal
Diagnostics
Published
2025-10-31
DOI
https://doi.org/10.3390/diagnostics15212777
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Photon-Counting Computed Tomography of the Paranasal Sinuses Improves Intraoperative Accuracy of Image-Guided Surgery

Thomas J. Vogl, Jan‐Erik Scholtz, Jonas Eckrich, Daniel Kuetting et al.
Diagnostics
Advanced X-ray and CT Imaging
article

Photon-Counting Computed Tomography of the Paranasal Sinuses Improves Intraoperative Accuracy of Image-Guided Surgery

Thomas J. Vogl, Jan‐Erik Scholtz, Jonas Eckrich, Daniel Kuetting, Amos Levi, Andreas Loth, Benjamin Philipp Ernst, Sebastian Strieth, Sven Becker, Matthias Frank Froelich, Stefanie Schliwa, Alexander Hertel, Friederike Bärhold, Tobias Albrecht, Iris Burck, Yannik C. Layer
article en

Abstract

<b>Background:</b> Computed tomography (CT)-based image-guided surgery (IGS) is of great importance in functional endoscopic sinus surgery (FESS) and requires IGS-specific imaging protocols to ensure high intraoperative accuracy. This study aimed to compare photon-counting CT (PCCT), dual-energy dual-source CT (DECT), and spectral detector CT (SDCT) of the paranasal sinuses with respect to image quality, IGS accuracy and radiation dose. <b>Methods:</b> A formalin-fixed cadaver skull was examined using PCCT, DECT and SDCT at 100 kV tube voltage with descending tube currents (mAs). The setup of electromagnetic IGS was evaluated using a visual analog scale. Accuracy was analyzed endoscopically using defined anatomical landmarks. Diagnostic image quality as well as bone and soft tissue noise were assessed qualitatively using a 5-point Likert scale and quantitatively by determination of signal-to-noise ratio. Radiation dose was evaluated using the dose length product. <b>Results:</b> While PCCT datasets could be registered and navigated accurately down to 10 mAs (1.5 mm error at 10 mAs), both DECT and SDCT exhibited significantly increased inaccuracies below 40 mAs (4.35/5.15 mm for DECT/SDCT at 25 mAs). Using PCCT therefore enabled a 45% radiation dose reduction at the minimally required dose length product using PCCT. Quantitative and qualitative image quality were superior for PCCT compared to DECT and SDCT. <b>Conclusions:</b> PCCT provides excellent accuracy of anatomical landmarks in IGS with superior image quality of the paranasal sinuses in low-mA scans and substantially reduced radiation exposure.

DiagnosticsVol. 15(21)
Goethe University Frankfurt (DE), University of Bonn (DE), Johannes Gutenberg University Mainz (DE), Heidelberg University (DE), University Hospital Heidelberg (DE), University Hospital Frankfurt (DE), University Medical Centre Mannheim (DE), University of Tübingen (DE)
Openalex Percentile: Top 69%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.