Resolving falsely elevated vancomycin levels in KIMS immunoassay due to monoclonal IgM interference

Given its narrow therapeutic index and risk of nephrotoxicity, vancomycin requires accurate concentration monitoring. While immunoassays such as the Kinetic Interaction of Microparticles in Solution (KIMS) are widely used for their speed and simplicity, they are susceptible to interference by paraproteins such as monoclonal immunoglobulins. Here, we present the first case of falsely elevated vancomycin concentrations measured by KIMS due to monoclonal IgM interference, indicated by a 'kinetic flag' error, and evaluated pre-analytical approaches to mitigate this interference. We applied three pre-analytical techniques to overcome this issue: dithiothreitol (DTT) treatment to reduce disulfide bonds in IgM, ultrafiltration with a 100-kDa cutoff to physically remove IgM molecules, and heterophilic blocking tubes (HBT) to neutralize nonspecific binding by heterophilic antibodies. Vancomycin concentrations were measured using both KIMS (Cobas c702, Roche Diagnostics) and Chemiluminescent Microparticle Immunoassay (CMIA, Architect i2000SR, Abbott Laboratories). In pre-vancomycin-dose samples, both DTT treatment and ultrafiltration resolved the 'kinetic flag' error and corrected vancomycin concentrations to below the lower limit of quantification (LLOQ). In post-vancomycin-dose samples, DTT reduced vancomycin concentration to 16.82 μg/mL (93.8% of the naïve concentration), while ultrafiltration reduced it to 9.22 μg/mL (51.4%). HBT treatment did not resolve the error, and falsely elevated results persisted. This study reports the first case of IgM paraprotein interfering with vancomycin measurement by KIMS, underlining the importance of recognizing monoclonal IgM as a potential source of interference. DTT treatment effectively corrected the interference, while ultrafiltration reduced vancomycin concentration, highlighting the need for careful interpretation.

Authors

Institutions

Publication Details

Journal
Scandinavian Journal of Clinical and Laboratory Investigation
Published
2025-10-03
DOI
https://doi.org/10.1080/00365513.2025.2582207
Primary Topic
Protein purification and stability
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Resolving falsely elevated vancomycin levels in KIMS immunoassay due to monoclonal IgM interference

Haeil Park, Go Eun Bae, Soo‐Youn Lee, Sangmi Kim
Scandinavian Journal of Clinical and Laboratory Investigation
Protein purification and stability
article

Resolving falsely elevated vancomycin levels in KIMS immunoassay due to monoclonal IgM interference

Haeil Park, Go Eun Bae, Soo‐Youn Lee, Sangmi Kim
article en

Abstract

Given its narrow therapeutic index and risk of nephrotoxicity, vancomycin requires accurate concentration monitoring. While immunoassays such as the Kinetic Interaction of Microparticles in Solution (KIMS) are widely used for their speed and simplicity, they are susceptible to interference by paraproteins such as monoclonal immunoglobulins. Here, we present the first case of falsely elevated vancomycin concentrations measured by KIMS due to monoclonal IgM interference, indicated by a 'kinetic flag' error, and evaluated pre-analytical approaches to mitigate this interference. We applied three pre-analytical techniques to overcome this issue: dithiothreitol (DTT) treatment to reduce disulfide bonds in IgM, ultrafiltration with a 100-kDa cutoff to physically remove IgM molecules, and heterophilic blocking tubes (HBT) to neutralize nonspecific binding by heterophilic antibodies. Vancomycin concentrations were measured using both KIMS (Cobas c702, Roche Diagnostics) and Chemiluminescent Microparticle Immunoassay (CMIA, Architect i2000SR, Abbott Laboratories). In pre-vancomycin-dose samples, both DTT treatment and ultrafiltration resolved the 'kinetic flag' error and corrected vancomycin concentrations to below the lower limit of quantification (LLOQ). In post-vancomycin-dose samples, DTT reduced vancomycin concentration to 16.82 μg/mL (93.8% of the naïve concentration), while ultrafiltration reduced it to 9.22 μg/mL (51.4%). HBT treatment did not resolve the error, and falsely elevated results persisted. This study reports the first case of IgM paraprotein interfering with vancomycin measurement by KIMS, underlining the importance of recognizing monoclonal IgM as a potential source of interference. DTT treatment effectively corrected the interference, while ultrafiltration reduced vancomycin concentration, highlighting the need for careful interpretation.

Scandinavian Journal of Clinical and Laboratory InvestigationVol. 85(7)
Samsung Medical Center (KR), Chosun University Hospital (KR), The Catholic University of Korea Bucheon St. Mary's Hospital (KR), Sungkyunkwan University (KR)
Openalex Percentile: Top 71%
Protein purification and stability
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.