Validation of ARTSENS Plus in Comparison to SphygmoCor XCEL for Assessing Arterial Stiffness During Pregnancy: A Cross-sectional Study

Abstract Background Arterial stiffness independently predicts cardiovascular mortality and morbidity. It is shown to be increased in vascular mediated conditions, such as preeclampsia and foetal growth restriction. ARTSENS Plus device assesses arterial stiffness and is validated in older populations. The study aimed to validate its use for measuring arterial stiffness in pregnant women by comparing it with SphygmoCor XCEL as a reference standard. Methods This cross-sectional study was conducted in two centres in the south-eastern region of India, recruiting 147 pregnant women. Arterial stiffness was assessed by the carotid–femoral pulse wave velocity (cfPWV) using both devices and validated according to the ‘2024 recommendations for validation of non-invasive arterial pulse wave velocity measurement devices.’ Bland–Altman plot and coefficient of variation of the test–retest reproducibility were calculated. Results The mean age of the included women was 28 ± 4 years, and their mean gestational age was 24.6 ± 9.5 weeks. The mean difference in cfPWV values obtained from SphygmoCor XCEL and ARTSENS Plus was 0.16 ± 0.54 m/s ( p = 0.12) with a measurement error of 0.85 m/s, indicating good accuracy of the ARTSENS plus device in assessing arterial stiffness. Both devices demonstrated good intra-observer reproducibility, with coefficients of variation of 2.53% for ARTSENS Plus and 3.48% for SphygmoCor XCEL. Conclusions ARTSENS Plus assesses arterial stiffness with good accuracy and intra-observer reproducibility when validated amongst pregnant women.

Authors

Institutions

Publication Details

Journal
Artery Research
Published
2025-11-03
DOI
https://doi.org/10.1007/s44200-025-00090-y
Primary Topic
Pregnancy and preeclampsia studies
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Validation of ARTSENS Plus in Comparison to SphygmoCor XCEL for Assessing Arterial Stiffness During Pregnancy: A Cross-sectional Study

Jayaraj Joseph, Dilip Kumar Maurya, Suresh Seshadri, P. M. Nabeel et al.
Artery Research
Pregnancy and preeclampsia studies
article

Validation of ARTSENS Plus in Comparison to SphygmoCor XCEL for Assessing Arterial Stiffness During Pregnancy: A Cross-sectional Study

Jayaraj Joseph, Dilip Kumar Maurya, Suresh Seshadri, P. M. Nabeel, Uma Ram, Anish Keepanasseril
article en

Abstract

Abstract Background Arterial stiffness independently predicts cardiovascular mortality and morbidity. It is shown to be increased in vascular mediated conditions, such as preeclampsia and foetal growth restriction. ARTSENS Plus device assesses arterial stiffness and is validated in older populations. The study aimed to validate its use for measuring arterial stiffness in pregnant women by comparing it with SphygmoCor XCEL as a reference standard. Methods This cross-sectional study was conducted in two centres in the south-eastern region of India, recruiting 147 pregnant women. Arterial stiffness was assessed by the carotid–femoral pulse wave velocity (cfPWV) using both devices and validated according to the ‘2024 recommendations for validation of non-invasive arterial pulse wave velocity measurement devices.’ Bland–Altman plot and coefficient of variation of the test–retest reproducibility were calculated. Results The mean age of the included women was 28 ± 4 years, and their mean gestational age was 24.6 ± 9.5 weeks. The mean difference in cfPWV values obtained from SphygmoCor XCEL and ARTSENS Plus was 0.16 ± 0.54 m/s ( p = 0.12) with a measurement error of 0.85 m/s, indicating good accuracy of the ARTSENS plus device in assessing arterial stiffness. Both devices demonstrated good intra-observer reproducibility, with coefficients of variation of 2.53% for ARTSENS Plus and 3.48% for SphygmoCor XCEL. Conclusions ARTSENS Plus assesses arterial stiffness with good accuracy and intra-observer reproducibility when validated amongst pregnant women.

Artery ResearchVol. 31(1)
Indian Institute of Technology Madras (IN), Healthcare Technology Innovation Centre (IN), Fetal Care Research Foundation (IN), Seethapathy Clinic and Hospital (IN), Jawaharlal Institute of Post Graduate Medical Education and Research (IN)
Indian Institute of Technology Madras, Ministry of Education, India, Division of Human Resource Development
Openalex Percentile: Top 60%
Pregnancy and preeclampsia studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.