Background: High-intensity ultrasound (HIU) catheter ablation produces deeper lesions than radiofrequency and may serve as a new modality for interventricular septal (IVS) reduction in hypertrophic cardiomyopathy (HCM). A proprietary “contact sensing” algorithm was developed to confirm orientation of the side-facing transducer toward the IVS. Hypothesis: Contact sensing will improve HIU ablation efficacy, resulting in larger and more consistent IVS lesions. Methods: A total of 31 healthy swine underwent HIU catheter (Fig A) ablation via femoral venous access targeting the RV side of the IVS. In 13 of the 31 swine, continuous transducer-tissue contact was monitored to confirm tissue contact (Fig B) and optimize catheter positioning prior to ablation. Each swine underwent 8-16 HIU sonications. Animals were euthanized immediately post-ablation, and lesion presence, dimensions, volume, and location were assessed by gross pathology. Lesion metrics are reported as mean ± standard deviation. Group comparisons (with vs. without contact sensing) were performed using unpaired, two-tailed homoscedastic t-tests for continuous variables and chi-squared tests for categorical outcomes. Results: Contact sensing increased lesion volume from 657 ± 901 to 2071 ± 1230 mm 3 (p=0.001) and lesion depth from 7 ± 7 mm to 15 ± 3 mm (p=0.001; Fig C,D). Spline analysis of gross pathology showed that the lesion area on the LV side of the IVS increased from 22 ± 23% without sensing to 55 ± 19% with sensing (p=0.0004; Fig C,D). Transient atrio-ventricular block occurred in both groups (Fig D) and resolved with catheter repositioning. Sustained AVB occurred in one swine without contact sensing. Conclusion: HIU transducer-tissue contact sensing significantly improved HIU ablation efficacy, producing larger, deeper, and more optimally located lesions for HCM septal reduction. Survival studies will confirm the degree of septal reduction.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2025-11-03
DOI
https://doi.org/10.1161/circ.152.suppl_3.4360977
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4360977: Tissue-Transducer Contact Sensing Improves High-Intensity Ultrasound Catheter Ablation in Healthy Swine

Babak Nazer, Stephanie Totten, Craig Morris, Jeff Thiel et al.
Circulation
Cardiac Arrhythmias and Treatments
article

Abstract 4360977: Tissue-Transducer Contact Sensing Improves High-Intensity Ultrasound Catheter Ablation in Healthy Swine

Babak Nazer, Stephanie Totten, Craig Morris, Jeff Thiel, David Giraud, Yak-Nam Wang, Naveed Arang, Phoebe Dijour
article en

Abstract

Background: High-intensity ultrasound (HIU) catheter ablation produces deeper lesions than radiofrequency and may serve as a new modality for interventricular septal (IVS) reduction in hypertrophic cardiomyopathy (HCM). A proprietary “contact sensing” algorithm was developed to confirm orientation of the side-facing transducer toward the IVS. Hypothesis: Contact sensing will improve HIU ablation efficacy, resulting in larger and more consistent IVS lesions. Methods: A total of 31 healthy swine underwent HIU catheter (Fig A) ablation via femoral venous access targeting the RV side of the IVS. In 13 of the 31 swine, continuous transducer-tissue contact was monitored to confirm tissue contact (Fig B) and optimize catheter positioning prior to ablation. Each swine underwent 8-16 HIU sonications. Animals were euthanized immediately post-ablation, and lesion presence, dimensions, volume, and location were assessed by gross pathology. Lesion metrics are reported as mean ± standard deviation. Group comparisons (with vs. without contact sensing) were performed using unpaired, two-tailed homoscedastic t-tests for continuous variables and chi-squared tests for categorical outcomes. Results: Contact sensing increased lesion volume from 657 ± 901 to 2071 ± 1230 mm 3 (p=0.001) and lesion depth from 7 ± 7 mm to 15 ± 3 mm (p=0.001; Fig C,D). Spline analysis of gross pathology showed that the lesion area on the LV side of the IVS increased from 22 ± 23% without sensing to 55 ± 19% with sensing (p=0.0004; Fig C,D). Transient atrio-ventricular block occurred in both groups (Fig D) and resolved with catheter repositioning. Sustained AVB occurred in one swine without contact sensing. Conclusion: HIU transducer-tissue contact sensing significantly improved HIU ablation efficacy, producing larger, deeper, and more optimally located lesions for HCM septal reduction. Survival studies will confirm the degree of septal reduction.

CirculationVol. 152(Suppl_3)
University of Washington (US), Coridea (United States) (US)
Openalex Percentile: Top 51%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.