Abstract 4342332: A Non-Contacting Blood Flow Sensor for Assessing Pediatric Vascular Graft Patency

Introduction: For infants born with single ventricle physiology, the initial palliation involves the implantation of a systemic to pulmonary vascular graft, typically a modified Blalock-Taussig shunt (mBTS) or a Sano shunt. However, irregular blood flow through the life-sustaining shunt can be life-threatening, and state-of-the-art monitoring approaches are either invasive or indirect. Alternative approaches for periodic monitoring of blood flow in a nonclinical setting are needed. Methods: A novel contactless blood flow sensor was developed for integration on the shunt exterior for early detection of flow irregularities. This sensing strategy avoids device-related thrombosis by combining the capacitively-coupled contactless conductivity detection (C4D) principle with time-of-flight measurements to quantify blood flow. Five metal rings are positioned along the shunt exterior, each capacitively-coupled to the conductive blood. The first and last electrodes are used to impose a sinusoidal driving signal and voltage differences are measured between adjacent pairs among the three central electrodes. Time varying conductivity can be tracked and by cross-correlation of the two recordings, in which the downstream measurement is a time-delayed copy of the upstream one, and knowing the precise spacing of the pairs, flow rate can be computed. Results: Sensor prototypes consisted of platinum rings fitted on polytetrafluoroethylene tubing. Capacitive coupling of the rings to blood was evaluated first. Response to varying solution conductivities was then characterized, demonstrating a linear response with the highest sensitivity at lower conductivities. Following successful sensor calibration measurements in test fluid, ex vivo flow measurements using bovine blood were collected at 30 and 50 mL/min. Based on the electrode locations, the delay for the flow rates was expected to be 100 and 70 ms, respectively. Peaks were observed at 100 ms and 60 ms, corresponding to 30 mL/min (0% error) and 42.9 mL/min (14.2% error), respectively. Errors were comparable to current “gold-standard” devices. Conclusions: We successfully demonstrated the feasibility of non-contact flowing monitoring in a format suitable for pediatric vascular shunts. This low power approach can support a fully implantable device to provide early warning of inadequate blood flow and inform timely intervention to reduce morbidity and mortality in infants with single ventricle physiology.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2025-11-03
DOI
https://doi.org/10.1161/circ.152.suppl_3.4342332
Citations
1
Primary Topic
Electrical and Bioimpedance Tomography
Type
article
Field-Weighted Citation Impact
0.73
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4342332: A Non-Contacting Blood Flow Sensor for Assessing Pediatric Vascular Graft Patency

Xuechun Wang, Trevor Hudson, R Y Chen, Ellis Meng et al.
1 citations
Circulation
Electrical and Bioimpedance Tomography
0.73
article

Abstract 4342332: A Non-Contacting Blood Flow Sensor for Assessing Pediatric Vascular Graft Patency

Xuechun Wang, Trevor Hudson, R Y Chen, Ellis Meng, Vishal Nigam, Jingjing Liang, Alanna Nowlen
article en
1 citations

Abstract

Introduction: For infants born with single ventricle physiology, the initial palliation involves the implantation of a systemic to pulmonary vascular graft, typically a modified Blalock-Taussig shunt (mBTS) or a Sano shunt. However, irregular blood flow through the life-sustaining shunt can be life-threatening, and state-of-the-art monitoring approaches are either invasive or indirect. Alternative approaches for periodic monitoring of blood flow in a nonclinical setting are needed. Methods: A novel contactless blood flow sensor was developed for integration on the shunt exterior for early detection of flow irregularities. This sensing strategy avoids device-related thrombosis by combining the capacitively-coupled contactless conductivity detection (C4D) principle with time-of-flight measurements to quantify blood flow. Five metal rings are positioned along the shunt exterior, each capacitively-coupled to the conductive blood. The first and last electrodes are used to impose a sinusoidal driving signal and voltage differences are measured between adjacent pairs among the three central electrodes. Time varying conductivity can be tracked and by cross-correlation of the two recordings, in which the downstream measurement is a time-delayed copy of the upstream one, and knowing the precise spacing of the pairs, flow rate can be computed. Results: Sensor prototypes consisted of platinum rings fitted on polytetrafluoroethylene tubing. Capacitive coupling of the rings to blood was evaluated first. Response to varying solution conductivities was then characterized, demonstrating a linear response with the highest sensitivity at lower conductivities. Following successful sensor calibration measurements in test fluid, ex vivo flow measurements using bovine blood were collected at 30 and 50 mL/min. Based on the electrode locations, the delay for the flow rates was expected to be 100 and 70 ms, respectively. Peaks were observed at 100 ms and 60 ms, corresponding to 30 mL/min (0% error) and 42.9 mL/min (14.2% error), respectively. Errors were comparable to current “gold-standard” devices. Conclusions: We successfully demonstrated the feasibility of non-contact flowing monitoring in a format suitable for pediatric vascular shunts. This low power approach can support a fully implantable device to provide early warning of inadequate blood flow and inform timely intervention to reduce morbidity and mortality in infants with single ventricle physiology.

CirculationVol. 152(Suppl_3)
University of Southern California (US), Seattle Children's Hospital (US), Harvard University Press (US)
Openalex Percentile: Top 22%
Electrical and Bioimpedance Tomography
0.73
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.