Abstract Sun105: Effects of a Flow Modulation Valve on Manual Ventilation Performance and Hemodynamics During Hemorrhagic Shock Resuscitation

Background: Hemorrhagic shock, precipitated by rapid blood loss, is highly lethal. Successful treatment requires rapid stabilization to prevent mortality. Ventilatory support may be required when accompanied by respiratory failure, and in austere prehospital settings is most likely to occur via manual bagging, which carries risks of barotrauma and hemodynamic compromise. Airway flow limiting devices can facilitate correct manual ventilation, but their utility in the hemorrhagic shock resuscitation setting is not well established. Objective: Evaluate the use of an airway flow limiting device on manual ventilation performance and hemodynamics in a preclinical model of hemorrhagic shock resuscitation. Methods: We sedated, anesthetized and intubated 20 mixed-breed Yorkshire swine, mean(SD) weight 53.1(4) kg. Animals were mechanically ventilated at 8mL/kg while undergoing instrumentation (pressure transducer in descending aorta via femoral cutdown for assessing mean arterial pressure (MAP) and systolic blood pressure (SBP)). Airway flow, pressure (P AW ) and tidal volume (V T ) were monitored continuously. We bled animals in 60cc draws until MAP <40mmHg, and then commenced manual bagging, with animals randomized to standard ventilation (“SHAM”) or flow modulated ventilation (“FM”; Sotair, SafeBVM, Inc). Respiratory rate was guided by a metronome, dynamically adjusted to maintain eucapnia. After 30 min, all withdrawn blood was returned and bagging continued for another 30 min. Performance and hemodynamic parameters were compared between groups and across shock&resuscitation phases (t-tests; ANOVA). a = 0.05 was the threshold for statistical significance. Results: Fourteen pigs survived initial hemorrhage until randomization (1 animal arrested during resuscitation) and initiation of bagging (FM, n = 7; SHAM, n = 7). Airway performance parameters differed between FM and SHAM (p<0.001). MAP and SBP did not differ between FM and SHAM groups but differed between phases (p<0.001). Conclusions: Bagged ventilations with a flow modulation device resulted in more favorable ventilation performance, but no differences in hemodynamics during hemorrhagic shock/resuscitation.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2025-11-03
DOI
https://doi.org/10.1161/circ.152.suppl_3.sun105
Primary Topic
Trauma, Hemostasis, Coagulopathy, Resuscitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract Sun105: Effects of a Flow Modulation Valve on Manual Ventilation Performance and Hemodynamics During Hemorrhagic Shock Resuscitation

Arthur S. Slutsky, Rameen Forghani, James J. Menegazzi, David D. Salcido et al.
Circulation
Trauma, Hemostasis, Coagulopathy, Resuscitation
article

Abstract Sun105: Effects of a Flow Modulation Valve on Manual Ventilation Performance and Hemodynamics During Hemorrhagic Shock Resuscitation

Arthur S. Slutsky, Rameen Forghani, James J. Menegazzi, David D. Salcido, Nina Lane, Jorge J. Gumucio
article en

Abstract

Background: Hemorrhagic shock, precipitated by rapid blood loss, is highly lethal. Successful treatment requires rapid stabilization to prevent mortality. Ventilatory support may be required when accompanied by respiratory failure, and in austere prehospital settings is most likely to occur via manual bagging, which carries risks of barotrauma and hemodynamic compromise. Airway flow limiting devices can facilitate correct manual ventilation, but their utility in the hemorrhagic shock resuscitation setting is not well established. Objective: Evaluate the use of an airway flow limiting device on manual ventilation performance and hemodynamics in a preclinical model of hemorrhagic shock resuscitation. Methods: We sedated, anesthetized and intubated 20 mixed-breed Yorkshire swine, mean(SD) weight 53.1(4) kg. Animals were mechanically ventilated at 8mL/kg while undergoing instrumentation (pressure transducer in descending aorta via femoral cutdown for assessing mean arterial pressure (MAP) and systolic blood pressure (SBP)). Airway flow, pressure (P AW ) and tidal volume (V T ) were monitored continuously. We bled animals in 60cc draws until MAP <40mmHg, and then commenced manual bagging, with animals randomized to standard ventilation (“SHAM”) or flow modulated ventilation (“FM”; Sotair, SafeBVM, Inc). Respiratory rate was guided by a metronome, dynamically adjusted to maintain eucapnia. After 30 min, all withdrawn blood was returned and bagging continued for another 30 min. Performance and hemodynamic parameters were compared between groups and across shock&resuscitation phases (t-tests; ANOVA). a = 0.05 was the threshold for statistical significance. Results: Fourteen pigs survived initial hemorrhage until randomization (1 animal arrested during resuscitation) and initiation of bagging (FM, n = 7; SHAM, n = 7). Airway performance parameters differed between FM and SHAM (p<0.001). MAP and SBP did not differ between FM and SHAM groups but differed between phases (p<0.001). Conclusions: Bagged ventilations with a flow modulation device resulted in more favorable ventilation performance, but no differences in hemodynamics during hemorrhagic shock/resuscitation.

CirculationVol. 152(Suppl_3)
University of Pittsburgh (US), University of Toronto (CA)
Openalex Percentile: Top 63%
Trauma, Hemostasis, Coagulopathy, Resuscitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.