Abstract 4363387: Porcine Model of Biventricular Pressure-Volume Loop Assessment during Impella 5.5 Support

Introduction: The Impella 5.5 is frequently employed to provide temporary mechanical circulatory support as bridge to recovery or durable therapy. Optimization of device management relies on understanding ventricular pressure-volume (PV) relationships. Aim: The aim of this study was to evaluate the biventricular (BiV) relationship with PV loop assessment during Impella 5.5 support in non-myopathic ventricles. Methods: Six domestic Yorkshire Pigs were used for this study. The Impella 5.5 (J&J MedTech, NJ) was inserted percutaneously under ultrasound and fluoroscopic guidance through either the carotid artery or aortic arch, using a 26Fr DrySeal Sheath (Gore, AZ). BiV PV loop conductance catheters (Sigma-M, CD Leycom, Hengelo, The Netherlands) were inserted via femoral venous access. The LV PV loop catheter is positioned via transeptal puncture and with intracardiac echocardiogram guidance. BiV PV loops are generated for each level of Impella 5.5 support. Results: With increasing levels of Impella 5.5 support, left ventricular (LV) PV loops shifted increasingly to the left and right ventricular (RV) loops to the right. Accordingly, LV end-diastolic volume (EDV) decreased (110±4 ml to 90±8 ml; p<0.001) and RV EDV increased (121±7 ml to 147±13 ml; p<0.001), from baseline to P9. Native LV systolic function decreased with increasing Impella 5.5 support, as demonstrated by a reduction in stroke volume (SV) (56±5 ml to 49±5 ml; p<0.001), cardiac output (CO) (3.4±0.8 ml/min to 2.0±0.3 ml/min; p<0.001), ejection fraction (EF) (51±4% to 44±6%; p<0.001), stroke work (SW) (4483±1023 ml*mmHg to 2383±719 ml*mmHg; p<0.001), and LV peak positive dP/dt (825 ± 184 mmHg/s to 523 ± 99 mmHg/s; p<0.001). Conversely, the native RV experienced preload-mediated increase in SV (65±7 ml to 86±20 ml; p<0.001), CO (3.7±0.8 ml/min to 4.3±0.9 ml/min; p<0.001), EF (53±4% to 58±11%; p<0.001), SW (918±143 ml*mmHg to 1529±365 ml*mmHg; p<0.001), and RV dP/dt (164±16 mmHg/s to 228±29 mmHg/s, p<0.001). Load independent contractility was assessed in the LV (1.64±0.33 mmHg/ml to 1.71±0.41 mmHg/ml; p=0.2) and RV (0.25±0.04 mmHg/ml to 0.29±0.05 mmHg/ml; p<0.001). Conclusions: This study highlights ventricular interdependence with Impella 5.5 support with opposing BiV trends. In a healthy heart, the RV can adequately compensate for increased preload offered by the device.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2025-11-03
DOI
https://doi.org/10.1161/circ.152.suppl_3.4363387
Primary Topic
Mechanical Circulatory Support Devices
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4363387: Porcine Model of Biventricular Pressure-Volume Loop Assessment during Impella 5.5 Support

Atta Behfar, Suraj Yalamuri, Stephen T. Tiffany, Andrew N. Rosenbaum et al.
Circulation
Mechanical Circulatory Support Devices
article

Abstract 4363387: Porcine Model of Biventricular Pressure-Volume Loop Assessment during Impella 5.5 Support

Atta Behfar, Suraj Yalamuri, Stephen T. Tiffany, Andrew N. Rosenbaum, Sara S. Inglis, Tyra A. Witt, Serena Joseph Rahme, Mary Nagel, Katrina Tollefsrud
article en

Abstract

Introduction: The Impella 5.5 is frequently employed to provide temporary mechanical circulatory support as bridge to recovery or durable therapy. Optimization of device management relies on understanding ventricular pressure-volume (PV) relationships. Aim: The aim of this study was to evaluate the biventricular (BiV) relationship with PV loop assessment during Impella 5.5 support in non-myopathic ventricles. Methods: Six domestic Yorkshire Pigs were used for this study. The Impella 5.5 (J&J MedTech, NJ) was inserted percutaneously under ultrasound and fluoroscopic guidance through either the carotid artery or aortic arch, using a 26Fr DrySeal Sheath (Gore, AZ). BiV PV loop conductance catheters (Sigma-M, CD Leycom, Hengelo, The Netherlands) were inserted via femoral venous access. The LV PV loop catheter is positioned via transeptal puncture and with intracardiac echocardiogram guidance. BiV PV loops are generated for each level of Impella 5.5 support. Results: With increasing levels of Impella 5.5 support, left ventricular (LV) PV loops shifted increasingly to the left and right ventricular (RV) loops to the right. Accordingly, LV end-diastolic volume (EDV) decreased (110±4 ml to 90±8 ml; p<0.001) and RV EDV increased (121±7 ml to 147±13 ml; p<0.001), from baseline to P9. Native LV systolic function decreased with increasing Impella 5.5 support, as demonstrated by a reduction in stroke volume (SV) (56±5 ml to 49±5 ml; p<0.001), cardiac output (CO) (3.4±0.8 ml/min to 2.0±0.3 ml/min; p<0.001), ejection fraction (EF) (51±4% to 44±6%; p<0.001), stroke work (SW) (4483±1023 ml*mmHg to 2383±719 ml*mmHg; p<0.001), and LV peak positive dP/dt (825 ± 184 mmHg/s to 523 ± 99 mmHg/s; p<0.001). Conversely, the native RV experienced preload-mediated increase in SV (65±7 ml to 86±20 ml; p<0.001), CO (3.7±0.8 ml/min to 4.3±0.9 ml/min; p<0.001), EF (53±4% to 58±11%; p<0.001), SW (918±143 ml*mmHg to 1529±365 ml*mmHg; p<0.001), and RV dP/dt (164±16 mmHg/s to 228±29 mmHg/s, p<0.001). Load independent contractility was assessed in the LV (1.64±0.33 mmHg/ml to 1.71±0.41 mmHg/ml; p=0.2) and RV (0.25±0.04 mmHg/ml to 0.29±0.05 mmHg/ml; p<0.001). Conclusions: This study highlights ventricular interdependence with Impella 5.5 support with opposing BiV trends. In a healthy heart, the RV can adequately compensate for increased preload offered by the device.

CirculationVol. 152(Suppl_3)
Mayo Clinic (US)
Openalex Percentile: Top 69%
Mechanical Circulatory Support Devices
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.