Background: Leadless pacemakers are an emerging technology offering a less invasive alternative to traditional transvenous systems. Bachmann bundle (BB) pacing is a novel atrial pacing strategy that may promote more physiologic atrial activation. However, data on the feasibility, safety, and procedural characteristics of leadless pacemaker implantation targeting the Bachmann bundle remains limited. Methods: This retrospective case series was conducted at a tertiary center in Iowa City, USA. We included consecutive patients who underwent attempted BB targeted leadless pacemaker implantation between April 2024 and May 2025. Eligibility for the procedure was determined by the implanting physician. Outcomes included procedural success, procedure time, fluoroscopy duration, radiation dose (mGy), device parameters, EKG parameters, and complications. BB targeting was guided by fluoroscopic landmarks. Descriptive statistics summarized baseline characteristics and outcomes. Results: 17 patients (mean age 67.2 ± 15.5 years; 53% male; 94% white) were included. Of those, 13 (76.5%) had successful BB region leadless pacemaker implantation. Most common indications were sinus node dysfunction (46.2%) and complete heart block (38.5%). The mean procedure time was 72.7 ± 35.1 minutes, fluoroscopy time 10 ± 6.1 minutes, and mean radiation dose was 56.6 ± 47.2 mGy. Initial device parameters (n=13) showed a mean capture threshold of 2.0 ± 0.8 V at 0.6 ± 0.3 ms, sensing amplitude of 1.9 ± 0.9 mV, and impedance of 308.5 ± 67.3 ohms. Follow-up device interrogation, available for 10 patients, showed a mean capture threshold decreased to 0.9 ± 0.5 V at 0.5 ± 0.3 ms, sensing amplitude increased to 3.5 ± 1.4 mV, and impedance decreased to 279 ± 35.7 ohms. P wave duration decreased post-implant (140.3 ± 21.5 ms) vs. pre-implant (154.9 ± 33.5 ms). No procedural complications were reported. Conclusions: Leadless pacemaker implantation targeting the Bachmann’s Bundle is feasible and safe, with a high success rate and encouraging early electrical outcomes. These findings support further investigation of physiologic atrial pacing using leadless technology.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2025-11-03
DOI
https://doi.org/10.1161/circ.152.suppl_3.4367732
Primary Topic
Cardiac pacing and defibrillation studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4367732: Leadless Pacemaker Implantation in the Region of Bachmann’s Bundle: Initial Results

Paari Dominic, Ashraf Alzahrani, Steven Bailin, Peter Farjo et al.
Circulation
Cardiac pacing and defibrillation studies
article

Abstract 4367732: Leadless Pacemaker Implantation in the Region of Bachmann’s Bundle: Initial Results

Paari Dominic, Ashraf Alzahrani, Steven Bailin, Peter Farjo, Leena Makhdum
article en

Abstract

Background: Leadless pacemakers are an emerging technology offering a less invasive alternative to traditional transvenous systems. Bachmann bundle (BB) pacing is a novel atrial pacing strategy that may promote more physiologic atrial activation. However, data on the feasibility, safety, and procedural characteristics of leadless pacemaker implantation targeting the Bachmann bundle remains limited. Methods: This retrospective case series was conducted at a tertiary center in Iowa City, USA. We included consecutive patients who underwent attempted BB targeted leadless pacemaker implantation between April 2024 and May 2025. Eligibility for the procedure was determined by the implanting physician. Outcomes included procedural success, procedure time, fluoroscopy duration, radiation dose (mGy), device parameters, EKG parameters, and complications. BB targeting was guided by fluoroscopic landmarks. Descriptive statistics summarized baseline characteristics and outcomes. Results: 17 patients (mean age 67.2 ± 15.5 years; 53% male; 94% white) were included. Of those, 13 (76.5%) had successful BB region leadless pacemaker implantation. Most common indications were sinus node dysfunction (46.2%) and complete heart block (38.5%). The mean procedure time was 72.7 ± 35.1 minutes, fluoroscopy time 10 ± 6.1 minutes, and mean radiation dose was 56.6 ± 47.2 mGy. Initial device parameters (n=13) showed a mean capture threshold of 2.0 ± 0.8 V at 0.6 ± 0.3 ms, sensing amplitude of 1.9 ± 0.9 mV, and impedance of 308.5 ± 67.3 ohms. Follow-up device interrogation, available for 10 patients, showed a mean capture threshold decreased to 0.9 ± 0.5 V at 0.5 ± 0.3 ms, sensing amplitude increased to 3.5 ± 1.4 mV, and impedance decreased to 279 ± 35.7 ohms. P wave duration decreased post-implant (140.3 ± 21.5 ms) vs. pre-implant (154.9 ± 33.5 ms). No procedural complications were reported. Conclusions: Leadless pacemaker implantation targeting the Bachmann’s Bundle is feasible and safe, with a high success rate and encouraging early electrical outcomes. These findings support further investigation of physiologic atrial pacing using leadless technology.

CirculationVol. 152(Suppl_3)
University of Iowa Health Care (US)
Openalex Percentile: Top 51%
Cardiac pacing and defibrillation studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.