Abstract 4366434: Non-Invasive Estimation of Pulse Wave Velocity in Pregnant Women Using a Wearable Smart Ring

Background: Pulse wave velocity (PWV) is a marker of arterial stiffness and cardiovascular risk. While conventionally measured using tonometry or oscillometric systems, recent advances suggest that photoplethysmography (PPG) derived features correlate with PWV, enabling estimation via wearables such as smart rings. Maternal cardiovascular adaptations during pregnancy, including systemic vasodilation and increased vascular compliance, follow a U-shaped trajectory. Compliance typically increases until the second trimester and then decreases toward term, a pattern observable through both PPG and PWV. This provides a test case for PPG-based PWV estimation. This study evaluates the accuracy of PPG-derived PWV (ePWV) and its ability to capture pregnancy-related physiological changes when monitored with a smart ring. Methods: Carotid-femoral PWV (cf-PWV) was measured simultaneously with Oura Ring PPG using the SphygmoCor XCEL system in a separate cohort of 300 healthy individuals. A model was trained on 200 participants using wearable-derived features and evaluated in the remaining 100. The model was then applied to an anomymised pregnancy dataset from 5,664 users of the third-generation Oura Ring who self-reported pregnancy. Night-time PPG signals were collected during sleep from one month before conception to six months postpartum. For each participant, a baseline was defined from the month preceding conception, and ePWV deltas were calculated as relative changes. Delta values were aligned to the observation period. Results: Validation of the ePWV algorithm against cf-PWV showed strong agreement (r = 0.76), with a mean difference of 0.210 m/s and SD of 0.999 m/s. ePWV followed the expected U-shaped pattern during pregnancy. A gradual decline was observed after conception, reaching a minimum near the end of the second trimester, followed by a steady rise into the third trimester. Additionally, ePWV peaked one week postpartum before returning to pre-pregnancy levels within six months. Conclusion: PPG-derived ePWV showed strong agreement with cf-PWV reference values enabling monitoring of arterial stiffness using wearable smart ring. In addition, the ePWV patterns captured pregnancy-related vascular adaptations, reproducing the U-shaped trajectory described in prior studies. These findings support the feasibility of wearables for scalable non-invasive monitoring of arterial stiffness in response to physiological adaptations.

Authors

Publication Details

Journal
Circulation
Published
2025-11-03
DOI
https://doi.org/10.1161/circ.152.suppl_3.4366434
Primary Topic
Non-Invasive Vital Sign Monitoring
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4366434: Non-Invasive Estimation of Pulse Wave Velocity in Pregnant Women Using a Wearable Smart Ring

Aleksi Rantanen, Pauli Ohukainen, Miska Valkonen
Circulation
Non-Invasive Vital Sign Monitoring
article

Abstract 4366434: Non-Invasive Estimation of Pulse Wave Velocity in Pregnant Women Using a Wearable Smart Ring

Aleksi Rantanen, Pauli Ohukainen, Miska Valkonen
article en

Abstract

Background: Pulse wave velocity (PWV) is a marker of arterial stiffness and cardiovascular risk. While conventionally measured using tonometry or oscillometric systems, recent advances suggest that photoplethysmography (PPG) derived features correlate with PWV, enabling estimation via wearables such as smart rings. Maternal cardiovascular adaptations during pregnancy, including systemic vasodilation and increased vascular compliance, follow a U-shaped trajectory. Compliance typically increases until the second trimester and then decreases toward term, a pattern observable through both PPG and PWV. This provides a test case for PPG-based PWV estimation. This study evaluates the accuracy of PPG-derived PWV (ePWV) and its ability to capture pregnancy-related physiological changes when monitored with a smart ring. Methods: Carotid-femoral PWV (cf-PWV) was measured simultaneously with Oura Ring PPG using the SphygmoCor XCEL system in a separate cohort of 300 healthy individuals. A model was trained on 200 participants using wearable-derived features and evaluated in the remaining 100. The model was then applied to an anomymised pregnancy dataset from 5,664 users of the third-generation Oura Ring who self-reported pregnancy. Night-time PPG signals were collected during sleep from one month before conception to six months postpartum. For each participant, a baseline was defined from the month preceding conception, and ePWV deltas were calculated as relative changes. Delta values were aligned to the observation period. Results: Validation of the ePWV algorithm against cf-PWV showed strong agreement (r = 0.76), with a mean difference of 0.210 m/s and SD of 0.999 m/s. ePWV followed the expected U-shaped pattern during pregnancy. A gradual decline was observed after conception, reaching a minimum near the end of the second trimester, followed by a steady rise into the third trimester. Additionally, ePWV peaked one week postpartum before returning to pre-pregnancy levels within six months. Conclusion: PPG-derived ePWV showed strong agreement with cf-PWV reference values enabling monitoring of arterial stiffness using wearable smart ring. In addition, the ePWV patterns captured pregnancy-related vascular adaptations, reproducing the U-shaped trajectory described in prior studies. These findings support the feasibility of wearables for scalable non-invasive monitoring of arterial stiffness in response to physiological adaptations.

CirculationVol. 152(Suppl_3)
Openalex Percentile: Top 69%
Non-Invasive Vital Sign Monitoring
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.