Abstract 4370550: Device-Related Thrombosis on Watchman Implant: A Case Study Highlighting the Importance of Hypercoagulable Workup

68-year-old female with history of non-valvular atrial fibrillation who was unable to tolerate anti-coagulation due to multiple episodes of GI bleeds. She underwent left atrial appendage occlusion with a WATCHMAN (Boston Scientific). A 45 day post watchman implant did not reveal a leak and showed appropriate occlusion. Three months post-implantation, routine transesophageal echocardiography (TEE) revealed a small thrombus on the device. Patient was placed back on Rivaroxaban for 3 months. Repeat TEE showed resolution of the thrombus, and was taken of AC. One month after her last TEE (7 months after implantation); The patient presented to the emergency department with acute onset of right-sided hemiparesis and aphasia. Initial work up ECG showed rate controlled atrial fibrillation, with no change from prior ECGs. No significant elevations in cardiac biomarkers. Computed tomography angiography showed an acute left middle cerebral artery occlusion, confirmed by MRI. Emergency thrombectomy was performed. Urgent TEE was obtained; showed a massive frond like thrombus that encompassed the left atrium and originated on the Watchman device. The thrombus originated at the site of the occluder device. Given the recurrence of thrombus, patient underwent a hypercoagulable workup which was consistent with Anti-phospholipid antibody syndrome(APS) (Lupus anticoagulant elevated). Patient was recommended to switch to warfarin as treatment for APS and ASA. Repeat TEE 6 months later show improvement in the size of the thrombus (1.33 cm by 0.76 cm). Thankfully, there were no more episodes for GI bleeds or strokes during this period of time. Device-related thrombosis risk: The incidence of device-related thrombosis following Watchman implantation is reported to be approximately 3-4%. While most cases respond to anticoagulation, some patients may develop recurrent thrombosis. It is important to understand the factors that contribute to thrombus formation. Importance of hypercoagulable workup: The patient's underlying antiphospholipid syndrome was not identified prior to device implantation. This case underscores the necessity of conducting a thorough hypercoagulable workup in patients considered for left atrial appendage occlusion devices. Regular follow-up with imaging studies is crucial for early detection and management of device-related complications. In this case, routine TEE at three months post-implantation allowed for timely detection and treatment of the initial thrombus.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2025-11-03
DOI
https://doi.org/10.1161/circ.152.suppl_3.4370550
Primary Topic
Atrial Fibrillation Management and Outcomes
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4370550: Device-Related Thrombosis on Watchman Implant: A Case Study Highlighting the Importance of Hypercoagulable Workup

Vikas Yellapu, Jovan Plamenac, Mihir Ankola, Eluwana Amaratunga et al.
Circulation
Atrial Fibrillation Management and Outcomes
article

Abstract 4370550: Device-Related Thrombosis on Watchman Implant: A Case Study Highlighting the Importance of Hypercoagulable Workup

Vikas Yellapu, Jovan Plamenac, Mihir Ankola, Eluwana Amaratunga, Purujit Thacker, Spandan Desai
article en

Abstract

68-year-old female with history of non-valvular atrial fibrillation who was unable to tolerate anti-coagulation due to multiple episodes of GI bleeds. She underwent left atrial appendage occlusion with a WATCHMAN (Boston Scientific). A 45 day post watchman implant did not reveal a leak and showed appropriate occlusion. Three months post-implantation, routine transesophageal echocardiography (TEE) revealed a small thrombus on the device. Patient was placed back on Rivaroxaban for 3 months. Repeat TEE showed resolution of the thrombus, and was taken of AC. One month after her last TEE (7 months after implantation); The patient presented to the emergency department with acute onset of right-sided hemiparesis and aphasia. Initial work up ECG showed rate controlled atrial fibrillation, with no change from prior ECGs. No significant elevations in cardiac biomarkers. Computed tomography angiography showed an acute left middle cerebral artery occlusion, confirmed by MRI. Emergency thrombectomy was performed. Urgent TEE was obtained; showed a massive frond like thrombus that encompassed the left atrium and originated on the Watchman device. The thrombus originated at the site of the occluder device. Given the recurrence of thrombus, patient underwent a hypercoagulable workup which was consistent with Anti-phospholipid antibody syndrome(APS) (Lupus anticoagulant elevated). Patient was recommended to switch to warfarin as treatment for APS and ASA. Repeat TEE 6 months later show improvement in the size of the thrombus (1.33 cm by 0.76 cm). Thankfully, there were no more episodes for GI bleeds or strokes during this period of time. Device-related thrombosis risk: The incidence of device-related thrombosis following Watchman implantation is reported to be approximately 3-4%. While most cases respond to anticoagulation, some patients may develop recurrent thrombosis. It is important to understand the factors that contribute to thrombus formation. Importance of hypercoagulable workup: The patient's underlying antiphospholipid syndrome was not identified prior to device implantation. This case underscores the necessity of conducting a thorough hypercoagulable workup in patients considered for left atrial appendage occlusion devices. Regular follow-up with imaging studies is crucial for early detection and management of device-related complications. In this case, routine TEE at three months post-implantation allowed for timely detection and treatment of the initial thrombus.

CirculationVol. 152(Suppl_3)
St. Luke's University Health Network (US), St. Luke's Hospital (US)
Openalex Percentile: Top 51%
Atrial Fibrillation Management and Outcomes
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.