Abstract Or103: Automated cardiac arrest detection incorporated into a wristband: validation in patients with induced ventricular fibrillation

Background: Survival from unwitnessed out-of-hospital cardiac arrest is poor, often due to delayed activation of the emergency medical chain. Wearable technology capable of automated cardiac arrest detection and alerting could trigger rapid medical assistance. In a previous study, we developed an algorithm for cardiac arrest detection based on photoplethysmography (PPG) data from patients with induced circulatory arrest, achieving 98% sensitivity. The next step is validation of the algorithm in patients with shockable cardiac arrest. Goal: To study the performance of the developed PPG-algorithm in patients with cardiac arrest based on induced ventricular fibrillation (VF). Methods: From the prospective multicenter study DETECT-1, we selected all adult patients who underwent short-lasting VF induction as standard practice during subcutaneous implantable cardioverter defibrillator (S-ICD). Patients wore a PPG-wristband (CardioWatch) during the entire procedure. A cardiac arrest event was defined as induced VF. Continuous electrocardiogram (ECG) and arterial blood pressure were monitored as a reference standard. The primary endpoint was the sensitivity for the detection of cardiac arrest, as assessed using the previously developed PPG-algorithm. This PPG-based algorithm detects the absence of pulsations based on signal amplitude and waveform characteristics. Results: In total, 14 patients were included with a median age of 46 years (IQR 33-54), of whom four where female. Fifteen VF inductions were performed, see Figure 1. The sensitivity for cardiac arrest detection was 100% (95% confidence interval [CI] 75%-100%). No false positive cardiac arrest alerts occurred in 15 hours of PPG data, resulting in a positive predictive value of 100% (95% CI 75%-100%). Return of PPG pulsations after the S-ICD shock were detected successfully in all patients. Conclusions: The PPG-algorithm for cardiac arrest detection performs excellent in the detection of induced VF. As a next step, the algorithm needs to be validated in non-shockable cardiac arrest and potential false positive alarms during daily life use need to be studied.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2025-11-03
DOI
https://doi.org/10.1161/circ.152.suppl_3.or103
Primary Topic
Cardiac Arrest and Resuscitation
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract Or103: Automated cardiac arrest detection incorporated into a wristband: validation in patients with induced ventricular fibrillation

Aysun Cetinyurek‐Yavuz, Eelko Ronner, Niels T B Scholte, Masih Mafi-Rad et al.
Circulation
Cardiac Arrest and Resuscitation
article

Abstract Or103: Automated cardiac arrest detection incorporated into a wristband: validation in patients with induced ventricular fibrillation

Aysun Cetinyurek‐Yavuz, Eelko Ronner, Niels T B Scholte, Masih Mafi-Rad, Sing‐Chien Yap, Marc A. Brouwer, Roos Edgar, Kevin Vernooy, Niels van Royen, Reinoud E. Knops, Judith Bonnes, Eric H. Boersma, Kambiz Ebrahimkheil, Rypko Beukema, Caroline W. Stegink Jansen, Peter Stas
article en

Abstract

Background: Survival from unwitnessed out-of-hospital cardiac arrest is poor, often due to delayed activation of the emergency medical chain. Wearable technology capable of automated cardiac arrest detection and alerting could trigger rapid medical assistance. In a previous study, we developed an algorithm for cardiac arrest detection based on photoplethysmography (PPG) data from patients with induced circulatory arrest, achieving 98% sensitivity. The next step is validation of the algorithm in patients with shockable cardiac arrest. Goal: To study the performance of the developed PPG-algorithm in patients with cardiac arrest based on induced ventricular fibrillation (VF). Methods: From the prospective multicenter study DETECT-1, we selected all adult patients who underwent short-lasting VF induction as standard practice during subcutaneous implantable cardioverter defibrillator (S-ICD). Patients wore a PPG-wristband (CardioWatch) during the entire procedure. A cardiac arrest event was defined as induced VF. Continuous electrocardiogram (ECG) and arterial blood pressure were monitored as a reference standard. The primary endpoint was the sensitivity for the detection of cardiac arrest, as assessed using the previously developed PPG-algorithm. This PPG-based algorithm detects the absence of pulsations based on signal amplitude and waveform characteristics. Results: In total, 14 patients were included with a median age of 46 years (IQR 33-54), of whom four where female. Fifteen VF inductions were performed, see Figure 1. The sensitivity for cardiac arrest detection was 100% (95% confidence interval [CI] 75%-100%). No false positive cardiac arrest alerts occurred in 15 hours of PPG data, resulting in a positive predictive value of 100% (95% CI 75%-100%). Return of PPG pulsations after the S-ICD shock were detected successfully in all patients. Conclusions: The PPG-algorithm for cardiac arrest detection performs excellent in the detection of induced VF. As a next step, the algorithm needs to be validated in non-shockable cardiac arrest and potential false positive alarms during daily life use need to be studied.

CirculationVol. 152(Suppl_3)
Radboud University Nijmegen (NL), Maastricht University Medical Centre (NL), Erasmus MC (NL), Radboud University Medical Center (NL), Reinier de Graaf Hospital (NL), Amsterdam University Medical Centers (NL), Erasmus University Rotterdam (NL)
Openalex Percentile: Top 53%
Cardiac Arrest and Resuscitation
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.