Abstract 4338116: FibriCheck Detection Capabilities for Atrial Fibrillation: A Multicenter Validation Study

Background: Atrial fibrillation (AF) is the most common arrhythmia worldwide and is associated with significant morbidity, mortality, and healthcare spending. Despite medical advances, AF remains underdiagnosed and undertreated, leading to preventable complications. FibriCheck © [Qompium NV, Hasselt, Belgium] is a medical analysis platform that uses an end-to-end algorithm to detect AF based on photoplethysmography (PPG) signals recorded on consumer smartphones. Purpose: The study aimed to validate FibriCheck in a large, multi-center and multi-national cohort on ten popular smartphone devices. Methods: A total of 236 patients were recruited from five independent, large academic centers in the United States and Europe. The FibriCheck system incorporates several convolutional neural networks to detect individual heartbeats, estimate average heart rate, and classify the rhythm based on PPG signals. Classification is verified by a FibriCheck technician. Classification performance was compared to the standard 12-lead electrocardiogram in the study population. Performance was assessed across clinical subgroups and smartphone devices. Results: FibriCheck demonstrated high overall accuracy and reliability in detecting AF without technician verification: accuracy 98.5% (95% CI: 98.0%-99.0%); sensitivity 96.3% (95% CI: 94.4%-97.7%); specificity 99.3% (95% CI: 98.8%-99.7%); positive predictive value 98.0% (95% CI: 96.5%-98.9%); negative predictive value 99.8% (95% CI: 99.6%-99.9%). Performance was not affected by smartphone device or the presence or absence of comorbid heart failure, vascular disease, hypertension, diabetes, or stroke. Sensitivity was reduced in those with darker skin tone and higher BMI, but this was mitigated by technician verification. Conclusions: The study confirms the high accuracy, sensitivity, and specificity of the FibriCheck algorithm in detecting AF across various smartphone models and clinical subgroups. These findings support the use of FibriCheck as a reliable, low-cost, and easily accessible tool for AF detection in a diverse patient population.

Authors

Institutions

Publication Details

Journal
Circulation
Published
2025-11-03
DOI
https://doi.org/10.1161/circ.152.suppl_3.4338116
Primary Topic
ECG Monitoring and Analysis
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Abstract 4338116: FibriCheck Detection Capabilities for Atrial Fibrillation: A Multicenter Validation Study

Dieter Nuyens, Stavros Stavrakis, Hein Heidbüchel, David J. Slotwiner et al.
Circulation
ECG Monitoring and Analysis
article

Abstract 4338116: FibriCheck Detection Capabilities for Atrial Fibrillation: A Multicenter Validation Study

Dieter Nuyens, Stavros Stavrakis, Hein Heidbüchel, David J. Slotwiner, John Sollee, Máximo Rivero‐Ayerza, Alexander Volodarskiy, Baljash Cheema, Christophe Buyck, Laurent Pison, Hugo Van Herendael, Lien Desteghe, James D. Thomas
article en

Abstract

Background: Atrial fibrillation (AF) is the most common arrhythmia worldwide and is associated with significant morbidity, mortality, and healthcare spending. Despite medical advances, AF remains underdiagnosed and undertreated, leading to preventable complications. FibriCheck © [Qompium NV, Hasselt, Belgium] is a medical analysis platform that uses an end-to-end algorithm to detect AF based on photoplethysmography (PPG) signals recorded on consumer smartphones. Purpose: The study aimed to validate FibriCheck in a large, multi-center and multi-national cohort on ten popular smartphone devices. Methods: A total of 236 patients were recruited from five independent, large academic centers in the United States and Europe. The FibriCheck system incorporates several convolutional neural networks to detect individual heartbeats, estimate average heart rate, and classify the rhythm based on PPG signals. Classification is verified by a FibriCheck technician. Classification performance was compared to the standard 12-lead electrocardiogram in the study population. Performance was assessed across clinical subgroups and smartphone devices. Results: FibriCheck demonstrated high overall accuracy and reliability in detecting AF without technician verification: accuracy 98.5% (95% CI: 98.0%-99.0%); sensitivity 96.3% (95% CI: 94.4%-97.7%); specificity 99.3% (95% CI: 98.8%-99.7%); positive predictive value 98.0% (95% CI: 96.5%-98.9%); negative predictive value 99.8% (95% CI: 99.6%-99.9%). Performance was not affected by smartphone device or the presence or absence of comorbid heart failure, vascular disease, hypertension, diabetes, or stroke. Sensitivity was reduced in those with darker skin tone and higher BMI, but this was mitigated by technician verification. Conclusions: The study confirms the high accuracy, sensitivity, and specificity of the FibriCheck algorithm in detecting AF across various smartphone models and clinical subgroups. These findings support the use of FibriCheck as a reliable, low-cost, and easily accessible tool for AF detection in a diverse patient population.

CirculationVol. 152(Suppl_3)
Northwestern University (US), NewYork–Presbyterian Hospital (US), Cornell University (US), New York Hospital Queens (US), Presbyterian Hospital (US), Antwerp University Hospital (BE), Ziekenhuis Oost-Limburg (BE), University of Oklahoma Health Sciences Center (US), Hasselt University (BE)
Openalex Percentile: Top 52%
ECG Monitoring and Analysis
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.