Abstract 4354762: Safety and feasibility of magnetic resonance imaging within the first week following transvenous pacing system implantation
Introduction: Patients with transvenous cardiac implantable electrical devices (CIEDs) including pacemakers are typically excluded from magnetic resonance imaging (MRI) examinations for at least 6-weeks following implant. However, many MRI scans are required on urgent or emergency basis, potentially resulting in suboptimal management for patients following a recent CIED implant. The absolute risks of adverse lead/generator events with early MRI have not been quantified. Specific transvenous pacing systems from one manufacturer (Abbott) do not mandate a routine 6-week post-implant restriction for MRI conditionality, although real-world human data investigating lead parameter stability with MRI early (<7 days) post-implant are lacking. Research Question: Can MRI be performed early (<7 days) after pacemaker implantation without safety events or significant changes to lead parameters? Methods: Patients undergoing implantation of Abbott Assurity™ MRI pacemakers with active-fixation MRI conditional leads (Abbott Tendril™ 52cm/58cm or UltiPace™ 65cm) with no contraindication to MRI underwent cardiac MRI within 7 days of implant. Patients were scanned at 1.5 or 3T in an asynchronous pacing mode. Device interrogation was performed before and after MRI. All patients provided written consent. Pre- and post-MRI lead parameter measurements were compared using the two one-sided t-Tests (TOST) equivalence test. Equivalence margins were prespecified: lead impedance ±30Ω, lead sensing ±0.5mV, lead threshold ±0.25V. Results: 20 patients (median age 77 (IQR 72-80) years, 75% male) were recruited. 65% received right ventricular pacing and 35% conduction system pacing. Median time from implant to MRI was 3 [IQR 1-5] days. All lead parameters were statistically equivalent pre- and post-MRI: Atrial lead impedance mean difference 5 Ω (90% CI -3 to +12), TOST p<0.001 Ventricular lead impedance -1 Ω (90% CI -9.2 to +7.2), TOST p<0.001 Sensed P wave 0.04 mV (90% CI -0.13 to +0.22), TOST p<0.001 Sensed R wave -0.06 mV (90% CI -0.28 to +0.17), TOST p=0.001 Atrial capture threshold -0.01 mV (90% CI -0.04 to +0.01), TOST p<0.001 Ventricular capture threshold -0.01 mV (90% CI -0.05 to +0.03), TOST p<0.001 Conclusion: MRI within 7 days of pacemaker implantation was safe, with no adverse events and no significant change in lead parameters. These findings provide real-world data to complement manufacturer guidance and support access to early MRI when indicated for patients with conditional pacemaker systems.
Authors
- Dimitrios Panagopoulos (ORCID: https://orcid.org/0000-0002-7918-5065)
- C Fielder Camm (ORCID: https://orcid.org/0000-0003-2122-4225)
- Julian O.M. Ormerod (ORCID: https://orcid.org/0000-0003-1612-8992)
- Oliver J. Rider (ORCID: https://orcid.org/0000-0003-1295-7769)
- Kim Rajappan (ORCID: https://orcid.org/0000-0001-8996-2983)
- Timothy R. Betts (ORCID: https://orcid.org/0000-0001-9063-9905)
- Pok‐Tin Tang (ORCID: https://orcid.org/0000-0003-2086-2743)
- Matthew Ginks (ORCID: https://orcid.org/0000-0002-4704-4532)
- Sheila Coelho Ramalho Vasconcelos Morais (ORCID: https://orcid.org/0000-0001-9831-0338)
- Benjamin Mothibe Bussmann (ORCID: https://orcid.org/0000-0003-3343-2590)
- Stefan Neubauer (ORCID: https://orcid.org/0000-0001-9017-5645)
- James Gamble (ORCID: https://orcid.org/0000-0001-9905-3337)
- Jonathan Raby (ORCID: https://orcid.org/0000-0002-2313-0479)
- Andrew Lewis (ORCID: https://orcid.org/0000-0002-3714-6711)
- Neil Herring (ORCID: https://orcid.org/0000-0001-8453-6133)
- Michala Pedersen
Institutions
- University of Oxford (GB)
- Oxford University Hospitals NHS Trust (GB)
Publication Details
- Journal
- Circulation
- Published
- 2025-11-03
- DOI
- https://doi.org/10.1161/circ.152.suppl_3.4354762
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00