Relationship between Phase Transition and Drug Release of Doxorubicin-Encapsulated Liposomes with Different Lipid Compositions Using Raman Spectroscopy

Liposomes have a more complex structure than conventional low-molecular-weight pharmaceuticals, so there is concern that quality evaluation items will be diverse and evaluation methods will be complex. Raman spectroscopy has recently attracted attention as a Process Analytical Technology in the pharmaceutical manufacturing, and its application is expected to expand to biopharmaceuticals and other drugs with complex manufacturing in the future. We have demonstrated that the combination of probe-type Raman spectrometer and partial least squares analysis enables real-time quantification of drug inclusion rate and drug release rate from liposomes, and is useful as a new quality assessment method for liposomes. In this study, we evaluated the phase transition of drug-encapsulated liposomes and the accompanying drug release by using Raman spectroscopy. Drug-encapsulated liposomes were prepared by preparing liposomes with different cholesterol (CHOL) ratios. The phase transition and drug release of liposomes were evaluated by using Raman spectroscopy. Raman spectroscopic measurements showed that the peak intensity of the phase transition was large in systems with low CHOL ratios, while it was low in systems with high CHOL ratios. In the drug release test, a decrease in the peak intensity of the drug-derived spectra over time was observed significantly in the low CHOL ratio system compared to the high CHOL ratio system which is supposed to release drug lower due to liquid-ordered phase, suggesting that the drug release property increased in the low CHOL system. Thus, Raman spectroscopy can be used to evaluate the phase transition and the associated drug release properties of liposomes.

Authors

Institutions

Publication Details

Journal
Biological and Pharmaceutical Bulletin
Published
2025-10-31
DOI
https://doi.org/10.1248/bpb.b25-00115
Citations
1
Primary Topic
Nanoparticle-Based Drug Delivery
Type
article
Field-Weighted Citation Impact
0.60

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Relationship between Phase Transition and Drug Release of Doxorubicin-Encapsulated Liposomes with Different Lipid Compositions Using Raman Spectroscopy

Ricardo Yamada, Keiji Hirota, Hiroshi Hisada, Kazuki Haneishi et al.
1 citations
Biological and Pharmaceutical Bulletin
Nanoparticle-Based Drug Delivery
0.60
article

Relationship between Phase Transition and Drug Release of Doxorubicin-Encapsulated Liposomes with Different Lipid Compositions Using Raman Spectroscopy

Ricardo Yamada, Keiji Hirota, Hiroshi Hisada, Kazuki Haneishi, Takumi Sato, Mika Yoshimura Fujii, Toshiro Fukami, Tatsuo Koide
article en
1 citations

Abstract

Liposomes have a more complex structure than conventional low-molecular-weight pharmaceuticals, so there is concern that quality evaluation items will be diverse and evaluation methods will be complex. Raman spectroscopy has recently attracted attention as a Process Analytical Technology in the pharmaceutical manufacturing, and its application is expected to expand to biopharmaceuticals and other drugs with complex manufacturing in the future. We have demonstrated that the combination of probe-type Raman spectrometer and partial least squares analysis enables real-time quantification of drug inclusion rate and drug release rate from liposomes, and is useful as a new quality assessment method for liposomes. In this study, we evaluated the phase transition of drug-encapsulated liposomes and the accompanying drug release by using Raman spectroscopy. Drug-encapsulated liposomes were prepared by preparing liposomes with different cholesterol (CHOL) ratios. The phase transition and drug release of liposomes were evaluated by using Raman spectroscopy. Raman spectroscopic measurements showed that the peak intensity of the phase transition was large in systems with low CHOL ratios, while it was low in systems with high CHOL ratios. In the drug release test, a decrease in the peak intensity of the drug-derived spectra over time was observed significantly in the low CHOL ratio system compared to the high CHOL ratio system which is supposed to release drug lower due to liquid-ordered phase, suggesting that the drug release property increased in the low CHOL system. Thus, Raman spectroscopy can be used to evaluate the phase transition and the associated drug release properties of liposomes.

Biological and Pharmaceutical BulletinVol. 48(11)
Meiji Pharmaceutical University (JP)
Chiba University
Openalex Percentile: Top 28%
Nanoparticle-Based Drug Delivery
0.60
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.