Automated dose-effect calibration curve for X-rays using the cytokinesis-block micronucleus assay

This article shows the development of a dose–effect calibration curve for X-ray exposures ranging from 0 to 4 Gy using the cytokinesis-block micronucleus assay and automated analysis—the first effort of its kind reported in Latin America. This work establishes a regional benchmark for high-throughput methodologies in cytogenetic biodosimetry, highlighting their potential to improve operational efficiency and reduce response times in radiological emergencies. Methods: Blood samples from six healthy donors were irradiated with X-rays at seven dose levels (0–4 Gy) using a calibrated 6 MV linear accelerator. Two blind samples (1.5 and 3 Gy) were included for validation. The CBMN assay was performed following IAEA protocols, DAPI-stained slides were analyzed using a AxioImager.Z2 automated microscope integrated with MetaSystems Metafer4 and the MNScoreX classifier software. A negative binomial regression model (NB1) was used for model fitting, accounting for overdispersion in micronucleus (MN) frequency. Results: Automated scoring of binucleated lymphocytes showed a dose-dependent increase in MN frequency. The fitted model followed a linear–quadratic relationship: Y = 0.0545 + 0.0448·D + 0.0145·D², with all coefficients statistically significant (p < 0.001). Dose estimates for blinded samples (1.5 and 3 Gy) matched the true doses within 95% confidence intervals, with all z-scores < |3|. Conclusions: The resulting linear–quadratic dose–response curve enabled accurate estimation of blinded sample doses, with all z-scores falling within acceptable fitness-for-purpose thresholds. These results underscore the value of combining automated microscopy with robust statistical modeling to achieve reliable dose assessment, particularly in high-throughput settings and radiological emergency scenarios.

Authors

Institutions

Publication Details

Journal
Brazilian Journal of Radiation Sciences
Published
2025-10-31
DOI
https://doi.org/10.15392/2319-0612.2025.2908
Primary Topic
Carcinogens and Genotoxicity Assessment
Type
article
Field-Weighted Citation Impact
0.00

Funders

Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Automated dose-effect calibration curve for X-rays using the cytokinesis-block micronucleus assay

Fabio Andrés Chaves-Campos, Fernando Ortíz Morales, Anthony Cordero-Ramírez, Jorge Ernesto González Mesa et al.
Brazilian Journal of Radiation Sciences
Carcinogens and Genotoxicity Assessment
article

Automated dose-effect calibration curve for X-rays using the cytokinesis-block micronucleus assay

Fabio Andrés Chaves-Campos, Fernando Ortíz Morales, Anthony Cordero-Ramírez, Jorge Ernesto González Mesa, Julián Alonso Gómez-Castro
article en

Abstract

This article shows the development of a dose–effect calibration curve for X-ray exposures ranging from 0 to 4 Gy using the cytokinesis-block micronucleus assay and automated analysis—the first effort of its kind reported in Latin America. This work establishes a regional benchmark for high-throughput methodologies in cytogenetic biodosimetry, highlighting their potential to improve operational efficiency and reduce response times in radiological emergencies. Methods: Blood samples from six healthy donors were irradiated with X-rays at seven dose levels (0–4 Gy) using a calibrated 6 MV linear accelerator. Two blind samples (1.5 and 3 Gy) were included for validation. The CBMN assay was performed following IAEA protocols, DAPI-stained slides were analyzed using a AxioImager.Z2 automated microscope integrated with MetaSystems Metafer4 and the MNScoreX classifier software. A negative binomial regression model (NB1) was used for model fitting, accounting for overdispersion in micronucleus (MN) frequency. Results: Automated scoring of binucleated lymphocytes showed a dose-dependent increase in MN frequency. The fitted model followed a linear–quadratic relationship: Y = 0.0545 + 0.0448·D + 0.0145·D², with all coefficients statistically significant (p < 0.001). Dose estimates for blinded samples (1.5 and 3 Gy) matched the true doses within 95% confidence intervals, with all z-scores < |3|. Conclusions: The resulting linear–quadratic dose–response curve enabled accurate estimation of blinded sample doses, with all z-scores falling within acceptable fitness-for-purpose thresholds. These results underscore the value of combining automated microscopy with robust statistical modeling to achieve reliable dose assessment, particularly in high-throughput settings and radiological emergency scenarios.

Brazilian Journal of Radiation SciencesVol. 13(4)
Universidad Nacional (CR), Universidad de Costa Rica (CR), Facultad Latinoamericana de Ciencias Costa Rica (CR), Instituto Nacional de Higiene, Epidemiología y Microbiología (CU), Latin University of Costa Rica (CR), Centro de Aplicaciones Tecnológicas y Desarrollo Nuclear (CU)
International Atomic Energy Agency, Universidad de Costa Rica
Openalex Percentile: Top 61%
Carcinogens and Genotoxicity Assessment
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.