Sub-0.5 mm Resolution PET Scanner with 3-Layer DOI Detectors for Rodent Neuroimaging

Spatial resolution is the most important parameter for preclinical positron emission tomography (PET) to visualize mouse brain function with high quantification accuracy. However, the spatial resolution of PET has been limited to over 0.5 mm, which causes a substantial partial volume effect especially for small mouse brain structures. In this study, we present the initial results of a mouse brain dedicated PET scanner that can achieve sub-0.5 mm resolution. The ring diameter and axial coverage of the PET scanner are 48 mm and 23.4 mm. To encode depth-of-interaction (DOI) information, 3-layers of lutetium yttrium oxyorthosilicate crystals were stacked in a staggered configuration and coupled to a 5×5 array of silicon photomultipliers having a pixel pitch of 2.4 mm. The crystal pitch and total thickness are 0.8 mm and 11 mm. The PET performance was characterized according to the National Electrical Manufacturers Association NU4-2008 standard. In vivo mouse brain imaging was carried out with <sup>18</sup>F-FITM and <sup>18</sup>F-FDG tracers. The average radial resolution from the center to 10 mm offset was 0.67±0.06 mm with filtered back projection. The 0.45 mm diameter rods were identified clearly with an iterative reconstruction algorithm. To the best of our knowledge, this is the first separate identification of the hypothalamus, amygdala, and cerebellar nuclei of mouse brain. The developed PET scanner achieved sub-0.5 mm resolution thereby visualizing small mouse brain structures with high quantification accuracy.

Authors

Institutions

Publication Details

Journal
IEEE Transactions on Medical Imaging
Published
2025-01-01
DOI
https://doi.org/10.1109/tmi.2025.3627815
Citations
1
Primary Topic
Medical Imaging Techniques and Applications
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Sub-0.5 mm Resolution PET Scanner with 3-Layer DOI Detectors for Rodent Neuroimaging

Makoto Higuchi, Hidekatsu Wakizaka, Hideaki Tashima, Han Gyu Kang et al.
1 citations
IEEE Transactions on Medical Imaging
Medical Imaging Techniques and Applications
1.36
article

Sub-0.5 mm Resolution PET Scanner with 3-Layer DOI Detectors for Rodent Neuroimaging

Makoto Higuchi, Hidekatsu Wakizaka, Hideaki Tashima, Han Gyu Kang, Taiga Yamaya
article en
1 citations

Abstract

Spatial resolution is the most important parameter for preclinical positron emission tomography (PET) to visualize mouse brain function with high quantification accuracy. However, the spatial resolution of PET has been limited to over 0.5 mm, which causes a substantial partial volume effect especially for small mouse brain structures. In this study, we present the initial results of a mouse brain dedicated PET scanner that can achieve sub-0.5 mm resolution. The ring diameter and axial coverage of the PET scanner are 48 mm and 23.4 mm. To encode depth-of-interaction (DOI) information, 3-layers of lutetium yttrium oxyorthosilicate crystals were stacked in a staggered configuration and coupled to a 5×5 array of silicon photomultipliers having a pixel pitch of 2.4 mm. The crystal pitch and total thickness are 0.8 mm and 11 mm. The PET performance was characterized according to the National Electrical Manufacturers Association NU4-2008 standard. In vivo mouse brain imaging was carried out with <sup>18</sup>F-FITM and <sup>18</sup>F-FDG tracers. The average radial resolution from the center to 10 mm offset was 0.67±0.06 mm with filtered back projection. The 0.45 mm diameter rods were identified clearly with an iterative reconstruction algorithm. To the best of our knowledge, this is the first separate identification of the hypothalamus, amygdala, and cerebellar nuclei of mouse brain. The developed PET scanner achieved sub-0.5 mm resolution thereby visualizing small mouse brain structures with high quantification accuracy.

IEEE Transactions on Medical ImagingVol. PP
National Institutes for Quantum Science and Technology (JP)
Openalex Percentile: Top 14%
Medical Imaging Techniques and Applications
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.