A Quality by Design Approach to Greener Ultrahigh‐Performance Liquid Chromatography Method Development for the Quantitative and Impurity Analysis of Favipiravir

ABSTRACT A robust, rapid, and environmentally conscious ultrahigh‐performance liquid chromatography (UHPLC) method was developed and validated for the quantification of favipiravir and its related substances—favipiravir‐D4 (internal standard) and favipiravir‐5‐hydroxy—in pharmaceutical formulations. Method development was guided by a quality‐by‐design (QbD) framework, utilizing response surface methodology to model the effects of critical variables—ethanol concentration, mobile phase pH, and flow rate—on chromatographic responses including retention time (RT), tailing factor (TF), and theoretical plates (TP). Optimal separation was achieved with 20% ethanol, 0.2 mL/min flow rate, and pH 2.8, delivering high resolution and efficiency. Validation was performed according to ICH Q2(R1) guidelines, demonstrating excellent linearity ( R 2 > 0.999), high sensitivity with LOQ at 0.05%, accuracy with recoveries averaging 99.2%, and precision with %RSD consistently below 2%. Robustness testing confirmed method reliability under intentional variations of critical parameters. Forced degradation studies—including acidic, alkaline, oxidative, thermal, photolytic, and hydrolytic stress—confirmed the stability‐indicating capability, with complete separation of favipiravir from its degradation products. The method was successfully applied to quantify favipiravir in a marketed formulation (Avigan), confirming its suitability for routine quality control and stability testing. Environmental assessments using Complementary Green Analytical Procedure Index (ComplexGAPI), Analytical GREEnness Metric (AGREE), AGREEprep, Analytical Eco‐Scale, and Blue Applicability Grade Index (BAGI) (scores > 0.70) demonstrated minimal reagent use, operational efficiency, and green analytical performance. Overall, the developed UHPLC method provides a sustainable, high‐throughput, and regulatory‐compliant platform for routine pharmaceutical analysis, impurity profiling, and stability studies, exemplifying the integration of green chemistry principles with advanced method optimization in modern analytical workflows.

Authors

Institutions

Publication Details

Journal
Separation Science Plus
Published
2025-10-31
DOI
https://doi.org/10.1002/sscp.70153
Citations
1
Primary Topic
Analytical Chemistry and Chromatography
Type
article
Field-Weighted Citation Impact
0.76
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

A Quality by Design Approach to Greener Ultrahigh‐Performance Liquid Chromatography Method Development for the Quantitative and Impurity Analysis of Favipiravir

Venkata Kanaka Srivani Maddala, Kumaraswamy Gandla, KiranKumar Chagarlamudi
1 citations
Separation Science Plus
Analytical Chemistry and Chromatography
0.76
article

A Quality by Design Approach to Greener Ultrahigh‐Performance Liquid Chromatography Method Development for the Quantitative and Impurity Analysis of Favipiravir

Venkata Kanaka Srivani Maddala, Kumaraswamy Gandla, KiranKumar Chagarlamudi
article en
1 citations

Abstract

ABSTRACT A robust, rapid, and environmentally conscious ultrahigh‐performance liquid chromatography (UHPLC) method was developed and validated for the quantification of favipiravir and its related substances—favipiravir‐D4 (internal standard) and favipiravir‐5‐hydroxy—in pharmaceutical formulations. Method development was guided by a quality‐by‐design (QbD) framework, utilizing response surface methodology to model the effects of critical variables—ethanol concentration, mobile phase pH, and flow rate—on chromatographic responses including retention time (RT), tailing factor (TF), and theoretical plates (TP). Optimal separation was achieved with 20% ethanol, 0.2 mL/min flow rate, and pH 2.8, delivering high resolution and efficiency. Validation was performed according to ICH Q2(R1) guidelines, demonstrating excellent linearity ( R 2 > 0.999), high sensitivity with LOQ at 0.05%, accuracy with recoveries averaging 99.2%, and precision with %RSD consistently below 2%. Robustness testing confirmed method reliability under intentional variations of critical parameters. Forced degradation studies—including acidic, alkaline, oxidative, thermal, photolytic, and hydrolytic stress—confirmed the stability‐indicating capability, with complete separation of favipiravir from its degradation products. The method was successfully applied to quantify favipiravir in a marketed formulation (Avigan), confirming its suitability for routine quality control and stability testing. Environmental assessments using Complementary Green Analytical Procedure Index (ComplexGAPI), Analytical GREEnness Metric (AGREE), AGREEprep, Analytical Eco‐Scale, and Blue Applicability Grade Index (BAGI) (scores > 0.70) demonstrated minimal reagent use, operational efficiency, and green analytical performance. Overall, the developed UHPLC method provides a sustainable, high‐throughput, and regulatory‐compliant platform for routine pharmaceutical analysis, impurity profiling, and stability studies, exemplifying the integration of green chemistry principles with advanced method optimization in modern analytical workflows.

Separation Science PlusVol. 8(11)
Chaitanya Bharathi Institute of Technology (IN), Vignan's Foundation for Science, Technology & Research (IN)
Openalex Percentile: Top 24%
Analytical Chemistry and Chromatography
0.76
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.