Monte Carlo investigation of Ge-doped silica fiber detectors for precision small field dosimetry

Accurate dosimetry in small radiation fields remains a critical challenge in modern radiotherapy due to high dose gradients and limited detector spatial resolution. This study investigates the dosimetric performance of Ge-doped silica fiber radioluminescence (RL) detectors for small field measurements through Monte Carlo simulations using the TOPAS tool. Reference data from IAEA database were utilized for implementation of small radiation fields and validation, by comparing experimental beam profile and depth dose measurements for various field sizes. Quality dependence factors (QDFs) were calculated based on the formalism of the IAEA-AAPM working group, considering a Varian Clinac iX linear accelerator with a 10×10 cm 2 reference field and various small fields of 0.5×0.5 cm 2 , 1×1 cm 2 , 2×2 cm 2 and 4×4 cm 2 . The study explored different fiber detector diameters from 125 μm to 2 mm to optimize spatial resolution for small field dosimetry. Simulated electron spectra in various situations were analyzed in both water and fiber volumes to understand reasons behind detector response variations. Notably, the detector diameter significantly influenced quality dependence, with competing effects between energy dependence and a field size dependent spectral discrepancy leading to an optimal balance at a 0.5 mm fiber diameter. Calculated QDFs aligned well within the range reported for other detector types, supporting the applicability of Ge-doped silica fiber RL detectors for precise dose measurements in small radiation fields, offering a potential alternative for advanced radiotherapy applications. • Monte Carlo simulations assess QDF in Ge-doped silica fiber dosimeters • Spectral differences in fiber volume and water affect dose measurement • Smaller fiber diameters show higher low-energy electron contribution • 0.5 mm diameter identified as optimal with minimum QDF correction needed • Results support fiber dosimeter customization for small-field applications

Authors

Institutions

Publication Details

Journal
Radiation Measurements
Published
2025-10-31
DOI
https://doi.org/10.1016/j.radmeas.2025.107552
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Monte Carlo investigation of Ge-doped silica fiber detectors for precision small field dosimetry

F. Moradi, H.T. Zubair, Amin Khodaei, H.A. Abdul-Rashid et al.
Radiation Measurements
Advanced Radiotherapy Techniques
article

Monte Carlo investigation of Ge-doped silica fiber detectors for precision small field dosimetry

F. Moradi, H.T. Zubair, Amin Khodaei, H.A. Abdul-Rashid, S.A. Ibrahim
article en

Abstract

Accurate dosimetry in small radiation fields remains a critical challenge in modern radiotherapy due to high dose gradients and limited detector spatial resolution. This study investigates the dosimetric performance of Ge-doped silica fiber radioluminescence (RL) detectors for small field measurements through Monte Carlo simulations using the TOPAS tool. Reference data from IAEA database were utilized for implementation of small radiation fields and validation, by comparing experimental beam profile and depth dose measurements for various field sizes. Quality dependence factors (QDFs) were calculated based on the formalism of the IAEA-AAPM working group, considering a Varian Clinac iX linear accelerator with a 10×10 cm 2 reference field and various small fields of 0.5×0.5 cm 2 , 1×1 cm 2 , 2×2 cm 2 and 4×4 cm 2 . The study explored different fiber detector diameters from 125 μm to 2 mm to optimize spatial resolution for small field dosimetry. Simulated electron spectra in various situations were analyzed in both water and fiber volumes to understand reasons behind detector response variations. Notably, the detector diameter significantly influenced quality dependence, with competing effects between energy dependence and a field size dependent spectral discrepancy leading to an optimal balance at a 0.5 mm fiber diameter. Calculated QDFs aligned well within the range reported for other detector types, supporting the applicability of Ge-doped silica fiber RL detectors for precise dose measurements in small radiation fields, offering a potential alternative for advanced radiotherapy applications. • Monte Carlo simulations assess QDF in Ge-doped silica fiber dosimeters • Spectral differences in fiber volume and water affect dose measurement • Smaller fiber diameters show higher low-energy electron contribution • 0.5 mm diameter identified as optimal with minimum QDF correction needed • Results support fiber dosimeter customization for small-field applications

Radiation MeasurementsVol. 190
TATI University College (MY), Multimedia University (MY), University of Malaya (MY)
Openalex Percentile: Top 58%
Advanced Radiotherapy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.