Intraoperative Repair of Pilot Balloon Inflation Line Damage: A Practical Solution to Tube Replacement

Intraoperative damage to the pilot balloon inflation line of an endotracheal tube (ETT) can result in cuff deflation, leading to ineffective ventilation and increased aspiration risk. This case report describes a practical repair technique that allowed continued saf e ventilation without tube replacement. A 70-year-old woman underwent elective coronary artery bypass grafting. During central venous catheter placement, the anesthesia machine displayed low etCO₂ and ventilator failure alarms. Diminished bilateral breath sounds were noted, and the pilot balloon was found severed. As the patient's hemodynamics remained stable, a 20G intravenous catheter steel needle t ip was inserted directly into the inflation line, carefully positioned to avoid further damage. The cuff was reinflated using a syringe, and appropriate pressure was maintained with a three-way stopcock and cuff manometer. The air leak resolved, and surgery proceeded without requiring tube exchange. Pilot balloon damage can be safely managed intraoperatively using readily available equipment. This practical approach may serve as a valuable alternative to endotracheal tube replacement, especially in cases involving a difficult airw ay. Familiarity with such repair methods can enhance airway safety during surgery.

Authors

Publication Details

Journal
Van Medical Journal
Published
2025-01-01
DOI
https://doi.org/10.5505/vmj.2025.46320
Primary Topic
Airway Management and Intubation Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Intraoperative Repair of Pilot Balloon Inflation Line Damage: A Practical Solution to Tube Replacement

Hacı Yusuf Güneş, Atakan Aydın
Van Medical Journal
Airway Management and Intubation Techniques
article

Intraoperative Repair of Pilot Balloon Inflation Line Damage: A Practical Solution to Tube Replacement

Hacı Yusuf Güneş, Atakan Aydın
article en

Abstract

Intraoperative damage to the pilot balloon inflation line of an endotracheal tube (ETT) can result in cuff deflation, leading to ineffective ventilation and increased aspiration risk. This case report describes a practical repair technique that allowed continued saf e ventilation without tube replacement. A 70-year-old woman underwent elective coronary artery bypass grafting. During central venous catheter placement, the anesthesia machine displayed low etCO₂ and ventilator failure alarms. Diminished bilateral breath sounds were noted, and the pilot balloon was found severed. As the patient's hemodynamics remained stable, a 20G intravenous catheter steel needle t ip was inserted directly into the inflation line, carefully positioned to avoid further damage. The cuff was reinflated using a syringe, and appropriate pressure was maintained with a three-way stopcock and cuff manometer. The air leak resolved, and surgery proceeded without requiring tube exchange. Pilot balloon damage can be safely managed intraoperatively using readily available equipment. This practical approach may serve as a valuable alternative to endotracheal tube replacement, especially in cases involving a difficult airw ay. Familiarity with such repair methods can enhance airway safety during surgery.

Van Medical JournalVol. 32(4)
Openalex Percentile: Top 51%
Airway Management and Intubation Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.