An Annular CMUT Array and Acquisition Strategy for Continuous Monitoring

In many monitoring scenarios, repeated and operator-independent assessments are needed. Wearable ultrasound technology has the potential to continuously provide the vital information traditionally obtained from conventional ultrasound scanners, such as in fetal monitoring for high-risk pregnancies. This work is an engineering study motivated by that setting. A 144-element annular capacitive micromachined ultrasonic transducer (CMUT) is hereby proposed for 3-D ultrasound imaging. The array is characterized by its compact size and cost-effectiveness, with a geometry and low-voltage operation that make it a candidate for future wearable integration. To enhance the imaging performance, we propose the utilization of a Fermat's spiral virtual source (VS) pattern for diverging wave transmission and conduct a performance comparison with other VS patterns and standard techniques, such as focused and plane waves. To facilitate this analysis, a simplified and versatile simulation framework, enhanced by GPU acceleration, has been developed. The validation of the simulation framework aligned closely with expected values (0.002 ≤ MAE ≤ 0.089). VSs following a Fermat's spiral led to a balanced outcome across metrics, outperforming focused wave transmissions for this specific aperture. The proposed transducer presents imaging limitations that could be improved in future developments, but it establishes a foundational framework for the design and fabrication of cost-effective, compact 2-D transducers suitable for 3-D ultrasound imaging, with potential for future integration into wearable devices.

Authors

Institutions

Publication Details

Journal
Sensors
Published
2025-10-29
DOI
https://doi.org/10.3390/s25216637
Citations
1
Primary Topic
Ultrasound Imaging and Elastography
Type
article
Field-Weighted Citation Impact
1.36
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

An Annular CMUT Array and Acquisition Strategy for Continuous Monitoring

Hans‐Martin Schwab, A.H.J. Immink, Vincent Henneken, Amir Gholampour et al.
1 citations
Sensors
Ultrasound Imaging and Elastography
1.36
article

An Annular CMUT Array and Acquisition Strategy for Continuous Monitoring

Hans‐Martin Schwab, A.H.J. Immink, Vincent Henneken, Amir Gholampour, Richard G. P. Lopata, Willem-Jan de Wijs, R. van Schaijk, María José Almario Escorcia
article en
1 citations

Abstract

In many monitoring scenarios, repeated and operator-independent assessments are needed. Wearable ultrasound technology has the potential to continuously provide the vital information traditionally obtained from conventional ultrasound scanners, such as in fetal monitoring for high-risk pregnancies. This work is an engineering study motivated by that setting. A 144-element annular capacitive micromachined ultrasonic transducer (CMUT) is hereby proposed for 3-D ultrasound imaging. The array is characterized by its compact size and cost-effectiveness, with a geometry and low-voltage operation that make it a candidate for future wearable integration. To enhance the imaging performance, we propose the utilization of a Fermat's spiral virtual source (VS) pattern for diverging wave transmission and conduct a performance comparison with other VS patterns and standard techniques, such as focused and plane waves. To facilitate this analysis, a simplified and versatile simulation framework, enhanced by GPU acceleration, has been developed. The validation of the simulation framework aligned closely with expected values (0.002 ≤ MAE ≤ 0.089). VSs following a Fermat's spiral led to a balanced outcome across metrics, outperforming focused wave transmissions for this specific aperture. The proposed transducer presents imaging limitations that could be improved in future developments, but it establishes a foundational framework for the design and fabrication of cost-effective, compact 2-D transducers suitable for 3-D ultrasound imaging, with potential for future integration into wearable devices.

SensorsVol. 25(21)
NXP (Netherlands) (NL), Philips (Finland) (FI), Philips (Netherlands) (NL), Eindhoven University of Technology (NL)
Openalex Percentile: Top 14%
Ultrasound Imaging and Elastography
1.36
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.