Reversed ventilatory gap capnometry in ventilated pediatric patients: a brief review for evidence of technical and physiological mechanisms

Capnometry measures end-tidal carbon dioxide (PEtCO2) in expired air and is standard practice for monitoring mechanical ventilation in pediatric critical care. PEtCO2 is often used as a non-invasive indicator for arterial CO2 (PaCO2), while under normal conditions the PaCO2-PEtCO2 difference is 2–5 mmHg due to imperfect matching of ventilation-perfusion (V/Q). However, in some pediatric patients, a reversed (negative) P(a-Et)CO2 gradient (ventilatory gap) can be observed, where PEtCO2 exceeds PaCO2. While this phenomenon is commonly linked to underlying technicalities, physiological explanations, including relatively high closing volumes in infants, poorly compliant lung tissue, and elevated mixed venous CO₂ (Pv̅CO₂) due to increased CO₂ production during critical illness, have also been described in the literature. However, current evidence for these mechanisms is limited, and further research is needed to clarify their roles. Possibly, awareness of both technical and physiological factors that may underlie a negative P(a–Et)CO₂ gradient can help clinicians interpret observations more cautiously and consider appropriate adjustments to mechanical ventilation strategies as needed.

Authors

Institutions

Publication Details

Journal
Intensive Care Medicine – Paediatric and Neonatal
Published
2025-10-29
DOI
https://doi.org/10.1007/s44253-025-00107-z
Primary Topic
Respiratory Support and Mechanisms
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reversed ventilatory gap capnometry in ventilated pediatric patients: a brief review for evidence of technical and physiological mechanisms

A. Reinout, Rozalinde Klein-Blommert, Dick G. Markhorst
Intensive Care Medicine – Paediatric and Neonatal
Respiratory Support and Mechanisms
article

Reversed ventilatory gap capnometry in ventilated pediatric patients: a brief review for evidence of technical and physiological mechanisms

A. Reinout, Rozalinde Klein-Blommert, Dick G. Markhorst
article en

Abstract

Capnometry measures end-tidal carbon dioxide (PEtCO2) in expired air and is standard practice for monitoring mechanical ventilation in pediatric critical care. PEtCO2 is often used as a non-invasive indicator for arterial CO2 (PaCO2), while under normal conditions the PaCO2-PEtCO2 difference is 2–5 mmHg due to imperfect matching of ventilation-perfusion (V/Q). However, in some pediatric patients, a reversed (negative) P(a-Et)CO2 gradient (ventilatory gap) can be observed, where PEtCO2 exceeds PaCO2. While this phenomenon is commonly linked to underlying technicalities, physiological explanations, including relatively high closing volumes in infants, poorly compliant lung tissue, and elevated mixed venous CO₂ (Pv̅CO₂) due to increased CO₂ production during critical illness, have also been described in the literature. However, current evidence for these mechanisms is limited, and further research is needed to clarify their roles. Possibly, awareness of both technical and physiological factors that may underlie a negative P(a–Et)CO₂ gradient can help clinicians interpret observations more cautiously and consider appropriate adjustments to mechanical ventilation strategies as needed.

Intensive Care Medicine – Paediatric and NeonatalVol. 3(1)
Amsterdam University Medical Centers (NL)
Openalex Percentile: Top 55%
Respiratory Support and Mechanisms
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.