ANALYTICAL PERFORMANCE OF THE DG READER NET IN IDENTIFYING CRITICAL RESULTS IN IMMUNOHEMATOLOGIC ASSAYS ANTI-HUMAN GLOBULIN PHASE–DEPENDENT IN THE GEL TECHNIQUE

Accurate identification of critical results in immunohematology is essential for transfusion safety in blood bank services. False-negative reactions in tests involving the anti-human globulin (AHG) phase can hinder the detection of clinically significant antigens and alloantibodies, compromising compatibility and transfusion safety. Therefore, we aim to investigate the analytical performance of the DG Reader NET (Grifols®) in detecting critical results in immunohematologic tests involving the AHG phase using gel centrifugation. Thus, analytical reproducibility between two automated readers, analytical efficacy and the agreement level between automated and manual readings were assessed. The assay included 500 hemagglutination reactions with various reactivity degrees (0, (+), 1+, and 2+; encompassing strong, weak, and doubtful reactions) in both direct and indirect antiglobulin tests. Direct reactions were centrifuged immediately. Indirect reactions were incubated at 37°C for 15 minutes. Following gel card centrifugation, the results were read by two blinded observers and by two validated autommated readers. Observers underwent previous calibration. Agreement was assessed using Cohen’s Kappa (K) coefficient or simple agreement and diagnostic performance via accuracy, sensitivity, specificity, and positive and negative predictive values (PPV and NPV). Statistical significance (α = 0.05) and confidence intervals (95%) were determined using Z-tests and bootstrap resampling (10,000 iterations). Chi-square tests or Fisher’s exact test were applied depending on cell frequencies. The automated reader showed an accuracy of 0.88 (0.85-0.91), sensitivity of 0.87 (0.84- 0.90), specificity of 0.89 (0.83-0.94), PPV of 0.96 (0.94-0.98), and NPV of 0.69 (0.62-0.76). Subgroup analysis revealed high specificity and NPV for negative samples (0.89; 0.84-0.94). For 1+ and 2+ reactions, sensitivity and PPV were 1.00 (0.97-1.00), demonstrating excellent discriminatory capacity at higher reactivity intensities (P < 0.001). However, for the (+) group, sensitivity was unsatisfactory (0.59; 0.51-0.68) with no significant difference, highlighting the reader’s lack to flag weakly positive or doubtful results. Regarding reproducibility, there was excellent agreement between the devices (K = 0.93; 0.90-0.95). By reactivity group, agreement was 0.94 (0.90-0.98) for negative, 0.81 (0.74-0.88) for (+), 0.98 (0.96-1.00) for 1+, and 0.96 (0.92-0.99) for 2+. These results indicate high automation reproducibility regardless of reactivity. Global agreement between reader and observer was also high (0.72; 0.67-0.77). For negative, 1+, and 2+ reactions, inter-method agreement exceeded 71% (63%-94%), while for (+) it was unsatisfactory (59%; 51%-68%). These findings demonstrate that although automation ensures high reproducibility and agreement for negative and positive reactions, there is a significant drop in agreement for weakly positive reactions (+). Despite the benefits of automation for analytical quality in blood banking, the data suggest that mandatory review criteria by qualified operators should be established for negative, weakly positive, and doubtful results generated by automated systems, in order to ensure the sensitivity required for immunohematological assays that are anti-human globulin–Dependent and guarantee safety in blood transfusion contexts.

Authors

Institutions

Publication Details

Journal
Hematology Transfusion and Cell Therapy
Published
2025-10-01
DOI
https://doi.org/10.1016/j.htct.2025.105622
Primary Topic
Blood groups and transfusion
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

ANALYTICAL PERFORMANCE OF THE DG READER NET IN IDENTIFYING CRITICAL RESULTS IN IMMUNOHEMATOLOGIC ASSAYS ANTI-HUMAN GLOBULIN PHASE–DEPENDENT IN THE GEL TECHNIQUE

Hebert Silveira, EJ Schörner, ACR de Moraes, David P. Siegel et al.
Hematology Transfusion and Cell Therapy
Blood groups and transfusion
article

ANALYTICAL PERFORMANCE OF THE DG READER NET IN IDENTIFYING CRITICAL RESULTS IN IMMUNOHEMATOLOGIC ASSAYS ANTI-HUMAN GLOBULIN PHASE–DEPENDENT IN THE GEL TECHNIQUE

Hebert Silveira, EJ Schörner, ACR de Moraes, David P. Siegel, Gabriel Fernandes Alves Jesus, Maria Passos Bianchini, Flávia Martinello
article en

Abstract

Accurate identification of critical results in immunohematology is essential for transfusion safety in blood bank services. False-negative reactions in tests involving the anti-human globulin (AHG) phase can hinder the detection of clinically significant antigens and alloantibodies, compromising compatibility and transfusion safety. Therefore, we aim to investigate the analytical performance of the DG Reader NET (Grifols®) in detecting critical results in immunohematologic tests involving the AHG phase using gel centrifugation. Thus, analytical reproducibility between two automated readers, analytical efficacy and the agreement level between automated and manual readings were assessed. The assay included 500 hemagglutination reactions with various reactivity degrees (0, (+), 1+, and 2+; encompassing strong, weak, and doubtful reactions) in both direct and indirect antiglobulin tests. Direct reactions were centrifuged immediately. Indirect reactions were incubated at 37°C for 15 minutes. Following gel card centrifugation, the results were read by two blinded observers and by two validated autommated readers. Observers underwent previous calibration. Agreement was assessed using Cohen’s Kappa (K) coefficient or simple agreement and diagnostic performance via accuracy, sensitivity, specificity, and positive and negative predictive values (PPV and NPV). Statistical significance (α = 0.05) and confidence intervals (95%) were determined using Z-tests and bootstrap resampling (10,000 iterations). Chi-square tests or Fisher’s exact test were applied depending on cell frequencies. The automated reader showed an accuracy of 0.88 (0.85-0.91), sensitivity of 0.87 (0.84- 0.90), specificity of 0.89 (0.83-0.94), PPV of 0.96 (0.94-0.98), and NPV of 0.69 (0.62-0.76). Subgroup analysis revealed high specificity and NPV for negative samples (0.89; 0.84-0.94). For 1+ and 2+ reactions, sensitivity and PPV were 1.00 (0.97-1.00), demonstrating excellent discriminatory capacity at higher reactivity intensities (P < 0.001). However, for the (+) group, sensitivity was unsatisfactory (0.59; 0.51-0.68) with no significant difference, highlighting the reader’s lack to flag weakly positive or doubtful results. Regarding reproducibility, there was excellent agreement between the devices (K = 0.93; 0.90-0.95). By reactivity group, agreement was 0.94 (0.90-0.98) for negative, 0.81 (0.74-0.88) for (+), 0.98 (0.96-1.00) for 1+, and 0.96 (0.92-0.99) for 2+. These results indicate high automation reproducibility regardless of reactivity. Global agreement between reader and observer was also high (0.72; 0.67-0.77). For negative, 1+, and 2+ reactions, inter-method agreement exceeded 71% (63%-94%), while for (+) it was unsatisfactory (59%; 51%-68%). These findings demonstrate that although automation ensures high reproducibility and agreement for negative and positive reactions, there is a significant drop in agreement for weakly positive reactions (+). Despite the benefits of automation for analytical quality in blood banking, the data suggest that mandatory review criteria by qualified operators should be established for negative, weakly positive, and doubtful results generated by automated systems, in order to ensure the sensitivity required for immunohematological assays that are anti-human globulin–Dependent and guarantee safety in blood transfusion contexts.

Hematology Transfusion and Cell TherapyVol. 47
Universidade Federal de Santa Catarina (BR), Fundação Centro de Hematologia e Hemoterapia de Minas Gerais (BR)
Openalex Percentile: Top 60%
Blood groups and transfusion
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.