Advanced noise-optimized virtual monochromatic imaging for improving the contrast of dual-energy CT pulmonary angiography

Abstract Background Pulmonary embolism (PE) is a serious condition requiring accurate imaging for timely diagnosis. While computed tomography pulmonary angiography (CTPA) is the standard tool, its quality may be limited by contrast timing or noise. Dual-energy CT enables virtual monochromatic imaging (VMI), which enhances contrast but suffers from high noise at low keV levels. Advanced VMI (AVMI) addresses this by optimizing noise while preserving contrast. The objective of this study was to evaluate the quantitative image quality of an advanced virtual monoenergetic imaging (AVMI) algorithm in comparison with the equivalent 120-kVp acquisition at dual-energy CT pulmonary angiography (DE-CTPA) in patients with suspected pulmonary embolism (PE). Methods We evaluated dual-source dual-energy CTPA examinations from 123 patients (58 men and 65 women; mean age, 54.2 ± 17.7 years; age range, 20–87 years) who had suspected pulmonary embolism (PE). Images were generated as AVMI series with energies ranging from 40- to 190-keV in 1-keV steps. Attenuation, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured in five different pulmonary arteries. AVMIs at 75-keV are considered equivalent to standard 120-kVp single energy CT (SECT) acquisition. The CNR and SNR were compared in AVMIs and 120-kVp equivalent images for each artery by repeated-measures analysis of variance (ANOVA) with post-hoc Bonferroni correction. Result The highest mean attenuation and CNR value of pulmonary arteries were observed in the 40-keV AVMI series, with significant CNR and SNR values compared to 120-kVp equivalent images ( p < 0.05). At 40-keV, energy level improvements of attenuation, SNR, and CNR by 225.8%, 31.0%, 51.2% were observed compared to 120-kVp equivalent images, respectively. Conclusion The advanced noise-optimized VMI series showed significantly higher SNR and CNR in comparison with 120-kVp equivalent images.

Authors

Institutions

Publication Details

Journal
The Egyptian Journal of Radiology and Nuclear Medicine
Published
2025-10-29
DOI
https://doi.org/10.1186/s43055-025-01609-8
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Advanced noise-optimized virtual monochromatic imaging for improving the contrast of dual-energy CT pulmonary angiography

Adnan Honardari, Hossein Ghadiri, Hadi Moradi, Mahsa Shakeri et al.
The Egyptian Journal of Radiology and Nuclear Medicine
Advanced X-ray and CT Imaging
article

Advanced noise-optimized virtual monochromatic imaging for improving the contrast of dual-energy CT pulmonary angiography

Adnan Honardari, Hossein Ghadiri, Hadi Moradi, Mahsa Shakeri, Ahmad Bitarafan‐Rajabi, Faezeh Shalbafzadeh
article en

Abstract

Abstract Background Pulmonary embolism (PE) is a serious condition requiring accurate imaging for timely diagnosis. While computed tomography pulmonary angiography (CTPA) is the standard tool, its quality may be limited by contrast timing or noise. Dual-energy CT enables virtual monochromatic imaging (VMI), which enhances contrast but suffers from high noise at low keV levels. Advanced VMI (AVMI) addresses this by optimizing noise while preserving contrast. The objective of this study was to evaluate the quantitative image quality of an advanced virtual monoenergetic imaging (AVMI) algorithm in comparison with the equivalent 120-kVp acquisition at dual-energy CT pulmonary angiography (DE-CTPA) in patients with suspected pulmonary embolism (PE). Methods We evaluated dual-source dual-energy CTPA examinations from 123 patients (58 men and 65 women; mean age, 54.2 ± 17.7 years; age range, 20–87 years) who had suspected pulmonary embolism (PE). Images were generated as AVMI series with energies ranging from 40- to 190-keV in 1-keV steps. Attenuation, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were measured in five different pulmonary arteries. AVMIs at 75-keV are considered equivalent to standard 120-kVp single energy CT (SECT) acquisition. The CNR and SNR were compared in AVMIs and 120-kVp equivalent images for each artery by repeated-measures analysis of variance (ANOVA) with post-hoc Bonferroni correction. Result The highest mean attenuation and CNR value of pulmonary arteries were observed in the 40-keV AVMI series, with significant CNR and SNR values compared to 120-kVp equivalent images ( p < 0.05). At 40-keV, energy level improvements of attenuation, SNR, and CNR by 225.8%, 31.0%, 51.2% were observed compared to 120-kVp equivalent images, respectively. Conclusion The advanced noise-optimized VMI series showed significantly higher SNR and CNR in comparison with 120-kVp equivalent images.

The Egyptian Journal of Radiology and Nuclear MedicineVol. 56(1)
Shaheed Rajaei Cardiovascular Medical and Research Center (IR), Tehran University of Medical Sciences (IR)
Openalex Percentile: Top 69%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.