Determining the Accuracy of Continuous Glucose Monitoring During Periods of Acute Glycemic Variability in Pregnancy

Background: The use of continuous glucose monitoring (CGM) in inpatient settings is highly attractive; it reduces the capillary glucose checks and enables remote monitoring. CGM systems have proven reliable in outpatient care; however, their reliability in hospitalized patients, particularly pregnant women, remains uncertain. Aim: To determine CGM accuracy and potential clinical risk of using CGM in pregnant women with type 1 diabetes mellitus (T1DM) or type-2 diabetes mellitus (T2DM) during acute hospitalization. Method: Prospective cohort study. CGM glucose was compared with paired capillary glucose readings as the reference. Concordance was assessed using mean absolute relative difference (MARD), the Food and Drug Administration standards for new integrated-CGM (iCGM) accuracy, and the diabetes technology society (DTS) error grid. Results: The dataset consisted of 1351 paired-readings from 26 women with T1DM and 11 women with T2DM. MARD was 12.9% overall, and 11.9% when the capillary glucose was 3.9–10.0 mmol/L. DTS error grid showed 81.6% of paired readings in zone A, 17.2% in zone B,1% in zone C and 0.1% in zone D. When capillary glucose was 3.9–10.0 mmol/L, 83.8% were in zone A, 15.5% in zone B, 0.6% in zone C, and 0.1% in zone D. Conclusions: These findings showed comparable accuracy in inpatient pregnant women to other inpatient groups in recent studies. CGM readings were highly correlated with the reference readings when capillary glucose was 3.9–10.0 mmol/L or when the CGM readings were 5.0–7.0 mmol/L. However, while CGM may guide inpatient antenatal diabetes care, critical management decisions (particularly around hypoglycemia) should be based on capillary glucose until greater concordance is established.

Authors

Institutions

Publication Details

Journal
Diabetes Technology and Obesity Medicine
Published
2025-10-01
DOI
https://doi.org/10.1177/29986702251389957
Primary Topic
Hyperglycemia and glycemic control in critically ill and hospitalized patients
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determining the Accuracy of Continuous Glucose Monitoring During Periods of Acute Glycemic Variability in Pregnancy

Ashleigh Stuart, Katie Wynne, Christopher W. Rowe, Naeel Mohammad
Diabetes Technology and Obesity Medicine
Hyperglycemia and glycemic control in critically ill and hospitalized patients
article

Determining the Accuracy of Continuous Glucose Monitoring During Periods of Acute Glycemic Variability in Pregnancy

Ashleigh Stuart, Katie Wynne, Christopher W. Rowe, Naeel Mohammad
article en

Abstract

Background: The use of continuous glucose monitoring (CGM) in inpatient settings is highly attractive; it reduces the capillary glucose checks and enables remote monitoring. CGM systems have proven reliable in outpatient care; however, their reliability in hospitalized patients, particularly pregnant women, remains uncertain. Aim: To determine CGM accuracy and potential clinical risk of using CGM in pregnant women with type 1 diabetes mellitus (T1DM) or type-2 diabetes mellitus (T2DM) during acute hospitalization. Method: Prospective cohort study. CGM glucose was compared with paired capillary glucose readings as the reference. Concordance was assessed using mean absolute relative difference (MARD), the Food and Drug Administration standards for new integrated-CGM (iCGM) accuracy, and the diabetes technology society (DTS) error grid. Results: The dataset consisted of 1351 paired-readings from 26 women with T1DM and 11 women with T2DM. MARD was 12.9% overall, and 11.9% when the capillary glucose was 3.9–10.0 mmol/L. DTS error grid showed 81.6% of paired readings in zone A, 17.2% in zone B,1% in zone C and 0.1% in zone D. When capillary glucose was 3.9–10.0 mmol/L, 83.8% were in zone A, 15.5% in zone B, 0.6% in zone C, and 0.1% in zone D. Conclusions: These findings showed comparable accuracy in inpatient pregnant women to other inpatient groups in recent studies. CGM readings were highly correlated with the reference readings when capillary glucose was 3.9–10.0 mmol/L or when the CGM readings were 5.0–7.0 mmol/L. However, while CGM may guide inpatient antenatal diabetes care, critical management decisions (particularly around hypoglycemia) should be based on capillary glucose until greater concordance is established.

Diabetes Technology and Obesity MedicineVol. 1(1)
John Hunter Hospital (AU), Hunter Medical Research Institute (AU), University of Newcastle Australia (AU)
Openalex Percentile: Top 57%
Hyperglycemia and glycemic control in critically ill and hospitalized patients
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.