Early Structural Deterioration of a Self-Expanding Transcatheter Aortic Valve Prosthesis Presenting With Severe Aortic Insufficiency: A Case Report

Transcatheter aortic valve replacement has become an established therapy for severe aortic stenosis, demonstrating prosthesis durability for at least five years (and up to eight years) in recent studies. Structural valve deterioration within 5 years is rare, particularly with self-expanding Evolut prostheses. We present an 88-year-old man with prior coronary artery bypass grafting and severe aortic stenosis who underwent transfemoral transcatheter aortic valve replacement with a 34 mm Evolut valve in 2019. Five years later, he developed progressive dyspnea and chest pain. Examination revealed a grade IV/VI decrescendo diastolic murmur and a markedly wide pulse pressure, raising concern for severe aortic regurgitation. Transesophageal echocardiography demonstrated moderate to severe aortic insufficiency. Coronary angiography revealed patent grafts, without new obstructive native coronary disease. He underwent surgical aortic valve replacement with an Inspiris Resilia bioprosthesis. This case underscores the possibility of rapid prosthetic valve failure and highlights the ongoing importance of physical examination in detecting severe regurgitation. It also contributes to the growing literature on early valve degeneration.

Authors

Institutions

Publication Details

Journal
Cureus
Published
2025-09-22
DOI
https://doi.org/10.7759/cureus.92971
Primary Topic
Cardiac Valve Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Early Structural Deterioration of a Self-Expanding Transcatheter Aortic Valve Prosthesis Presenting With Severe Aortic Insufficiency: A Case Report

Pranali R Dave, Melkon Hacobian
Cureus
Cardiac Valve Diseases and Treatments
article

Early Structural Deterioration of a Self-Expanding Transcatheter Aortic Valve Prosthesis Presenting With Severe Aortic Insufficiency: A Case Report

Pranali R Dave, Melkon Hacobian
article en

Abstract

Transcatheter aortic valve replacement has become an established therapy for severe aortic stenosis, demonstrating prosthesis durability for at least five years (and up to eight years) in recent studies. Structural valve deterioration within 5 years is rare, particularly with self-expanding Evolut prostheses. We present an 88-year-old man with prior coronary artery bypass grafting and severe aortic stenosis who underwent transfemoral transcatheter aortic valve replacement with a 34 mm Evolut valve in 2019. Five years later, he developed progressive dyspnea and chest pain. Examination revealed a grade IV/VI decrescendo diastolic murmur and a markedly wide pulse pressure, raising concern for severe aortic regurgitation. Transesophageal echocardiography demonstrated moderate to severe aortic insufficiency. Coronary angiography revealed patent grafts, without new obstructive native coronary disease. He underwent surgical aortic valve replacement with an Inspiris Resilia bioprosthesis. This case underscores the possibility of rapid prosthetic valve failure and highlights the ongoing importance of physical examination in detecting severe regurgitation. It also contributes to the growing literature on early valve degeneration.

CureusVol. 17(9)
University of California, Los Angeles (US), Arrowhead Regional Medical Center (US)
Openalex Percentile: Top 53%
Cardiac Valve Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Early Structural Deterioration of a Self-Expanding Transcatheter Aortic Valve Prosthesis Presenting With Severe Aortic Insufficiency: A Case Report — Pranali R Dave, Melkon Hacobian · Cureus (2025) | TGRS Research Map | TGRS