<i>Objective.</i>To design and build a motorized iris collimator suitable for a novel low-cost kilovoltage dual-robot radiotherapy device and test its reliability and dosimetric capabilities.<i>Approach.</i>A 12-leaflet motorized iris collimator was designed and built weighing 4.8 kg, with 1 mm thick tungsten leaflets, a beam size range of 0.1 cm to 16.5 cm, and a leaflet speed up to 20 cm/s. Dosimetric tests were performed with EBT4 Gafchromic film at depths of 0 to 10 cm in a Solid Water phantom at a source to surface distance of 38 cm using a 225 kVp beam with 2 mm of Al filtration. A Monte Carlo (MC) model of the experimental setup was developed. Beam size was determined by full width at half maximum (FWHM) and penumbra values were calculated for both film and MC. The beam size was measured experimentally using digital radiographs acquired for sizes between 5 and 150 mm to evaluate the robustness of the system.<i>Main results.</i>The collimator demonstrated sharp profiles with penumbra values ≤1.2 mm at 2 cm depth. 2D dose distributions from film measurements found dose leakage within the designed 2% as well as a single offset leaflet as a result of machining tolerance. MC matched film measurements with surface central dose differences ≤1.1%, a beam penumbra match within 0.08 mm, and a FWHM match within 1.1 mm. The linearity and repeatability showed that beam sizes can be set with an accuracy of ≤1.25 mm for beams ≥1 cm.<i>Significance.</i>This work showcases a motorized iris collimator designed for a novel low-cost radiotherapy device, capable of delivering sharp dose distributions. An MC model was successfully developed and matched to film measurements, allowing for accurate simulation of radiotherapy delivery.

Authors

Institutions

Publication Details

Journal
Biomedical Physics & Engineering Express
Published
2025-10-28
DOI
https://doi.org/10.1088/2057-1976/ae183e
Citations
2
Primary Topic
Advanced Radiotherapy Techniques
Type
article
Field-Weighted Citation Impact
3.77
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Development of a motorized iris collimator for kilovoltage x-ray radiotherapy

Nathan Clements, Magdalena Bazalova‐Carter, Nicolas Braam, Lane Aaron Braun et al.
2 citations
Biomedical Physics & Engineering Express
Advanced Radiotherapy Techniques
3.77
article

Development of a motorized iris collimator for kilovoltage x-ray radiotherapy

Nathan Clements, Magdalena Bazalova‐Carter, Nicolas Braam, Lane Aaron Braun, A. Godfrey, Michael L. Winn, Rolf Seuster, M Weil, J. Crosby Chapman, J Atkinson, Sandhya Rottoo, Olivia Masella, William Canthal
article en
2 citations

Abstract

<i>Objective.</i>To design and build a motorized iris collimator suitable for a novel low-cost kilovoltage dual-robot radiotherapy device and test its reliability and dosimetric capabilities.<i>Approach.</i>A 12-leaflet motorized iris collimator was designed and built weighing 4.8 kg, with 1 mm thick tungsten leaflets, a beam size range of 0.1 cm to 16.5 cm, and a leaflet speed up to 20 cm/s. Dosimetric tests were performed with EBT4 Gafchromic film at depths of 0 to 10 cm in a Solid Water phantom at a source to surface distance of 38 cm using a 225 kVp beam with 2 mm of Al filtration. A Monte Carlo (MC) model of the experimental setup was developed. Beam size was determined by full width at half maximum (FWHM) and penumbra values were calculated for both film and MC. The beam size was measured experimentally using digital radiographs acquired for sizes between 5 and 150 mm to evaluate the robustness of the system.<i>Main results.</i>The collimator demonstrated sharp profiles with penumbra values ≤1.2 mm at 2 cm depth. 2D dose distributions from film measurements found dose leakage within the designed 2% as well as a single offset leaflet as a result of machining tolerance. MC matched film measurements with surface central dose differences ≤1.1%, a beam penumbra match within 0.08 mm, and a FWHM match within 1.1 mm. The linearity and repeatability showed that beam sizes can be set with an accuracy of ≤1.25 mm for beams ≥1 cm.<i>Significance.</i>This work showcases a motorized iris collimator designed for a novel low-cost radiotherapy device, capable of delivering sharp dose distributions. An MC model was successfully developed and matched to film measurements, allowing for accurate simulation of radiotherapy delivery.

Biomedical Physics & Engineering ExpressVol. 11(6)
University of Victoria (CA), Ambient Micro (United States) (US), McGill University (CA)
Openalex Percentile: Top 6%
Advanced Radiotherapy Techniques
3.77
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.