Investigation of compact multi-layer ionization chamber detector designs for quality control of CBCT-based synthetic CTs for adaptive proton therapy

<i>Objective.</i>To define an optimal compact multi-layer ionization chamber (MLIC) detector design for proton radiography (PR) acquisition as quality control (QC) of cone-beam CT (CBCT)-based synthetic-CTs (sCTs) in the context of adaptive proton therapy (APT).<i>Approach.</i>In 10 patients, 51 PR measurements were simulated. To replicate measured IDDs (=IDDsMeas), noise based on actual PR measurements was added to the raw IDDs (=IDDsRaw). IDDsMeaswere discretized to 1-6 mm and limited to 30-50 mm to mimic MLIC detector designs with 30 different ionization chamber (IC) pitch and detector length combinations. The influence of noise and space/limited IDD information on the ability to determine range shifts was assessed by comparing IDDRawto IDDMeas. To assess how well different MLIC detector designs identify/detect sCT uncertainties, 43 error scenarios were generated by modifying CT calibration curves of muscle/fat/bone-like density tissues. The range shifts were calculated by fitting nominal IDDs to IDDsMeasof each error scenario using least-square. The normalized comparison of expected and measured range shifts is reported as normalized relative range shifts, ideally equaling zero.<i>Main results.</i>Each MLIC detector design was investigated based on 506 IDDRaw, with 100 different noise contributions each, resulting in 50600 IDDMeas. The relative range shift and relative normalized range shift increased with a larger IC pitch. An increased detector length allowed for a larger IC pitch. Detector lengths<35 mm proved insufficient for sampling enough relevant IDD information.<i>Significance.</i>This study demonstrates that range shifts due to density errors between measured and simulated IDDs can be identified using a compact MLIC detector. A normalized relative range shift uncertainty detection of⩽0.51%can be verified with a minimum 45 mm detector length and maximum 3 mm IC pitch. The obtained results suggest that a compact MLIC detector can be used to verify the integrity of CBCT-based sCTs, enabling APT based on sCTs.

Authors

Institutions

Publication Details

Journal
Physics in Medicine and Biology
Published
2025-10-28
DOI
https://doi.org/10.1088/1361-6560/ae1882
Primary Topic
Radiation Therapy and Dosimetry
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Investigation of compact multi-layer ionization chamber detector designs for quality control of CBCT-based synthetic CTs for adaptive proton therapy

Måns Lundberg, Artūrs Meijers, Antje Knopf, Kevin Souris et al.
Physics in Medicine and Biology
Radiation Therapy and Dosimetry
article

Investigation of compact multi-layer ionization chamber detector designs for quality control of CBCT-based synthetic CTs for adaptive proton therapy

Måns Lundberg, Artūrs Meijers, Antje Knopf, Kevin Souris, Antony Lomax
article en

Abstract

<i>Objective.</i>To define an optimal compact multi-layer ionization chamber (MLIC) detector design for proton radiography (PR) acquisition as quality control (QC) of cone-beam CT (CBCT)-based synthetic-CTs (sCTs) in the context of adaptive proton therapy (APT).<i>Approach.</i>In 10 patients, 51 PR measurements were simulated. To replicate measured IDDs (=IDDsMeas), noise based on actual PR measurements was added to the raw IDDs (=IDDsRaw). IDDsMeaswere discretized to 1-6 mm and limited to 30-50 mm to mimic MLIC detector designs with 30 different ionization chamber (IC) pitch and detector length combinations. The influence of noise and space/limited IDD information on the ability to determine range shifts was assessed by comparing IDDRawto IDDMeas. To assess how well different MLIC detector designs identify/detect sCT uncertainties, 43 error scenarios were generated by modifying CT calibration curves of muscle/fat/bone-like density tissues. The range shifts were calculated by fitting nominal IDDs to IDDsMeasof each error scenario using least-square. The normalized comparison of expected and measured range shifts is reported as normalized relative range shifts, ideally equaling zero.<i>Main results.</i>Each MLIC detector design was investigated based on 506 IDDRaw, with 100 different noise contributions each, resulting in 50600 IDDMeas. The relative range shift and relative normalized range shift increased with a larger IC pitch. An increased detector length allowed for a larger IC pitch. Detector lengths<35 mm proved insufficient for sampling enough relevant IDD information.<i>Significance.</i>This study demonstrates that range shifts due to density errors between measured and simulated IDDs can be identified using a compact MLIC detector. A normalized relative range shift uncertainty detection of⩽0.51%can be verified with a minimum 45 mm detector length and maximum 3 mm IC pitch. The obtained results suggest that a compact MLIC detector can be used to verify the integrity of CBCT-based sCTs, enabling APT based on sCTs.

Physics in Medicine and BiologyVol. 70(23)
University of Bern (CH), Bern University of Applied Sciences (CH), University Hospital of Bern (CH), Paul Scherrer Institute (CH), ETH Zurich (CH), Ion Beam Applications (Belgium) (BE), Cyclotron (Netherlands) (NL), University Hospital Carl Gustav Carus (DE), Technische Universität Dresden (DE)
Openalex Percentile: Top 55%
Radiation Therapy and Dosimetry
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.