Theoretical Study Synchronized Reverse Scan Collision-Induced Dissociation in Digital Linear Ion Trap

The effectiveness of collision-induced dissociation (CID) in ion trap mass spectrometry (ITMS) is limited by a low-mass cutoff and weak fragmentation yields. Theoretically, the <i>q</i> value is optimized to balance the fractional product ion mass range with adequate energy deposition to improve fragment ion detection in the CID process; however, many promising technologies still depend on the traditional sinusoidal waveform-driven IT. Additionally, traditional CID-based multistage mass spectrometry (MS<sup>n</sup>) experiments on ITMS rely on complex and time-consuming "tuning" to optimize CID for a particular ion. The digital ion trap (DIT) has a very promising application field in MS<sup>n</sup> analysis, because of its many unique features. Herein, we conducted a theoretical and experimental investigation of a developed synchronized reverse scan-CID (SRS-CID) using a digital linear ion trap. Specifically, (1) simulations and experiments demonstrated that in the SRS-CID, ions were sequentially scanned from high to low <i>m</i>/<i>z</i> value via the resonance excitation point (<i>q</i><sub>excitation</sub>), producing multiple fragment ions without the need to know the <i>m</i>/<i>z</i> value or complex radiofrequency (rf) tuning of each product ion. The simulations demonstrated that the heating rate in the SRS-CID could reach 0.022 eV/μs. The experiments demonstrated that the optimal reverse scan speed was -0.053 ns/step. (2) We preliminary increased the period by a fixed value (<i>T</i><sub>step</sub>) to control <i>q</i><sub>excitation</sub> to study the molecule fragmentation approach. Different mass spectra were obtained by controlling <i>t</i><sub>excitation</sub> with a fixed <i>T</i><sub>step</sub>. (3) This paper introduces the phase space method to study the motion trajectories of precursor ions and daughter ions. The calculations used and the entire program were uploaded to GitHub. (4) Changing the duty cycle to advantageously shift <i>q</i><sub>excitation</sub> improved the heating rate (0.033 eV/μs) in SRS-CID. Overall, we demonstrated the effectiveness of the developed SRS-CID technique in fragment ion analysis via theoretical derivation, simulation, and experimentation. Furthermore, DIT mass spectrometry was advantageous in tandem mass spectrometry analysis by facilitating modulation of the driving rf period.

Authors

Institutions

Publication Details

Journal
Journal of the American Society for Mass Spectrometry
Published
2025-10-28
DOI
https://doi.org/10.1021/jasms.5c00277
Primary Topic
Mass Spectrometry Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Theoretical Study Synchronized Reverse Scan Collision-Induced Dissociation in Digital Linear Ion Trap

Fuxing Xu, Weimin Wang, Li Ding, Chuan‐Fan Ding et al.
Journal of the American Society for Mass Spectrometry
Mass Spectrometry Techniques and Applications
article

Theoretical Study Synchronized Reverse Scan Collision-Induced Dissociation in Digital Linear Ion Trap

Fuxing Xu, Weimin Wang, Li Ding, Chuan‐Fan Ding, Z. Q. Xie
article en

Abstract

The effectiveness of collision-induced dissociation (CID) in ion trap mass spectrometry (ITMS) is limited by a low-mass cutoff and weak fragmentation yields. Theoretically, the <i>q</i> value is optimized to balance the fractional product ion mass range with adequate energy deposition to improve fragment ion detection in the CID process; however, many promising technologies still depend on the traditional sinusoidal waveform-driven IT. Additionally, traditional CID-based multistage mass spectrometry (MS<sup>n</sup>) experiments on ITMS rely on complex and time-consuming "tuning" to optimize CID for a particular ion. The digital ion trap (DIT) has a very promising application field in MS<sup>n</sup> analysis, because of its many unique features. Herein, we conducted a theoretical and experimental investigation of a developed synchronized reverse scan-CID (SRS-CID) using a digital linear ion trap. Specifically, (1) simulations and experiments demonstrated that in the SRS-CID, ions were sequentially scanned from high to low <i>m</i>/<i>z</i> value via the resonance excitation point (<i>q</i><sub>excitation</sub>), producing multiple fragment ions without the need to know the <i>m</i>/<i>z</i> value or complex radiofrequency (rf) tuning of each product ion. The simulations demonstrated that the heating rate in the SRS-CID could reach 0.022 eV/μs. The experiments demonstrated that the optimal reverse scan speed was -0.053 ns/step. (2) We preliminary increased the period by a fixed value (<i>T</i><sub>step</sub>) to control <i>q</i><sub>excitation</sub> to study the molecule fragmentation approach. Different mass spectra were obtained by controlling <i>t</i><sub>excitation</sub> with a fixed <i>T</i><sub>step</sub>. (3) This paper introduces the phase space method to study the motion trajectories of precursor ions and daughter ions. The calculations used and the entire program were uploaded to GitHub. (4) Changing the duty cycle to advantageously shift <i>q</i><sub>excitation</sub> improved the heating rate (0.033 eV/μs) in SRS-CID. Overall, we demonstrated the effectiveness of the developed SRS-CID technique in fragment ion analysis via theoretical derivation, simulation, and experimentation. Furthermore, DIT mass spectrometry was advantageous in tandem mass spectrometry analysis by facilitating modulation of the driving rf period.

Journal of the American Society for Mass SpectrometryVol. 36(12)
Ningbo University (CN), IONICS Mass Spectrometry (Canada) (CA)
Openalex Percentile: Top 71%
Mass Spectrometry Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.