In Vivo 3D Liver Imaging at 7 T Using kT‐Point pTx Pulses and a 32‐Tx‐Channel Whole‐Body Radiofrequency Coil Array

At 7 T, a main challenge is the flip angle variations arising from the spatially inhomogeneous B<sub>1</sub> <sup>+</sup> transmit profiles of the RF coils. To address this problem, static pTx with local pTx body arrays has been used. However, for larger organs, such as the liver, local arrays provide insufficient homogeneity, and often dynamic pTx is needed. In this study, the benefits of a 32-Tx-channel remote array are shown. Relative channel-wise 3D B<sub>1</sub> <sup>+</sup> maps are acquired in free breathing in 11:38 min with a radial phase encoding (RPE) acquisition scheme in five healthy volunteers at 7 T. For each subject, an ROI is drawn manually covering the whole liver. The necessary number of k<sub>T</sub>-point pTx pulses is analyzed by comparing L-curves that depict the tradeoff between FA homogeneity and RF power for a large range of regularization parameters. For all subjects, no FA dropouts remain, and CV values under 17% can be achieved with magnitude and phase shimming. Two to three k<sub>T</sub>-points showed a good tradeoff between FA homogeneity and RF power. To analyze different channel numbers, a fixed phase between the channel-wise B<sub>1</sub> <sup>+</sup> maps is set to achieve 8, 16, or 20 B<sub>1</sub> <sup>+</sup> maps for shimming. The higher channel number showed superior shimming performance for static and dynamic pTx. Comparing 32 versus 8 individual channels, a 1.3- to 1.6-fold improvement in homogeneity can be found for the same RF power using phase shimming. With 32 Tx channels, static pTx shim and two k<sub>T</sub>-points were applied in vivo in free breathing with a 3D RPE GRE sequence with 1.4-mm isotropic resolution. The in vivo signal magnitudes of the small FA acquisitions showed good agreement with the predicted relative B<sub>1</sub> <sup>+</sup> maps while neglecting B<sub>0</sub> nonuniformity. In summary, this study shows the advantages of using a remote 32-Tx channel array for in vivo liver imaging at 7 T.

Authors

Institutions

Publication Details

Journal
NMR in Biomedicine
Published
2025-10-28
DOI
https://doi.org/10.1002/nbm.70170
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

In Vivo 3D Liver Imaging at 7 T Using kT‐Point pTx Pulses and a 32‐Tx‐Channel Whole‐Body Radiofrequency Coil Array

Stephan Orzada, Sebastian Schmitter, Thomas M. Fiedler, Simon Schmidt et al.
NMR in Biomedicine
Advanced MRI Techniques and Applications
article

In Vivo 3D Liver Imaging at 7 T Using kT‐Point pTx Pulses and a 32‐Tx‐Channel Whole‐Body Radiofrequency Coil Array

Stephan Orzada, Sebastian Schmitter, Thomas M. Fiedler, Simon Schmidt, Mark E. Ladd, Harald H. Quick, Johannes A. Grimm, Sebastian Dietrich, Armin Michael Nagel, Christoph Stefan Aigner, Constantin Schorling
article en

Abstract

At 7 T, a main challenge is the flip angle variations arising from the spatially inhomogeneous B<sub>1</sub> <sup>+</sup> transmit profiles of the RF coils. To address this problem, static pTx with local pTx body arrays has been used. However, for larger organs, such as the liver, local arrays provide insufficient homogeneity, and often dynamic pTx is needed. In this study, the benefits of a 32-Tx-channel remote array are shown. Relative channel-wise 3D B<sub>1</sub> <sup>+</sup> maps are acquired in free breathing in 11:38 min with a radial phase encoding (RPE) acquisition scheme in five healthy volunteers at 7 T. For each subject, an ROI is drawn manually covering the whole liver. The necessary number of k<sub>T</sub>-point pTx pulses is analyzed by comparing L-curves that depict the tradeoff between FA homogeneity and RF power for a large range of regularization parameters. For all subjects, no FA dropouts remain, and CV values under 17% can be achieved with magnitude and phase shimming. Two to three k<sub>T</sub>-points showed a good tradeoff between FA homogeneity and RF power. To analyze different channel numbers, a fixed phase between the channel-wise B<sub>1</sub> <sup>+</sup> maps is set to achieve 8, 16, or 20 B<sub>1</sub> <sup>+</sup> maps for shimming. The higher channel number showed superior shimming performance for static and dynamic pTx. Comparing 32 versus 8 individual channels, a 1.3- to 1.6-fold improvement in homogeneity can be found for the same RF power using phase shimming. With 32 Tx channels, static pTx shim and two k<sub>T</sub>-points were applied in vivo in free breathing with a 3D RPE GRE sequence with 1.4-mm isotropic resolution. The in vivo signal magnitudes of the small FA acquisitions showed good agreement with the predicted relative B<sub>1</sub> <sup>+</sup> maps while neglecting B<sub>0</sub> nonuniformity. In summary, this study shows the advantages of using a remote 32-Tx channel array for in vivo liver imaging at 7 T.

NMR in BiomedicineVol. 38(12)
Erwin L. Hahn Institute for Magnetic Resonance Imaging (DE), Physikalisch-Technische Bundesanstalt (DE), University of Minnesota (US), German Cancer Research Center (DE), Friedrich-Alexander-Universität Erlangen-Nürnberg (DE), Heidelberg University (DE), Essen University Hospital (DE), Max Planck Institute for Human Development (DE), Resonance Research (United States) (US)
Openalex Percentile: Top 54%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.