Optimizing dosimetric leaf gap parameters for radiation therapy: A comparative study using three diode‐array verification tools and an ionization chamber
DLG differences between IC and diode array dosimeters resulted in approximately 1% dose discrepancies, which are within the clinically acceptable range. This finding indicates that ArcCHECK, Delta4, and SRS MapCHECK can determine DLG values comparable to those obtained with IC.