Integrated Parallel Beam WDS/EDS Analysis: Benefits and Limitations for Geoscience Research

The performance of an integrated parallel beam wavelength-dispersive/energy-dispersive microanalysis system mounted on a field emission scanning electron microscope has been compared with that of a conventional electron microprobe, using the same reference samples and as close as possible operational conditions. Tests on standards of known composition reveal internally consistent results for the former, and no obvious advantage of wavelength-dispersive over energy-dispersive spectroscopy has emerged for major elements without peak overlaps. Rather, because an energy-dispersive spectrometer can acquire all energies of the X-ray spectrum concurrently, conveying a lower electron dose per atom, it has the advantage of being faster and milder to minerals. In all cases, however, element migration may be significant, especially for monovalent, large ionic radius elements such as K and Na.

Authors

Institutions

Publication Details

Journal
Microscopy and Microanalysis
Published
2025-09-03
DOI
https://doi.org/10.1093/mam/ozaf101
Primary Topic
Electron and X-Ray Spectroscopy Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Integrated Parallel Beam WDS/EDS Analysis: Benefits and Limitations for Geoscience Research

Giancarlo Capitani, Michael Abratis, Stefano Zanchetta, Roberto Conconi et al.
Microscopy and Microanalysis
Electron and X-Ray Spectroscopy Techniques
article

Integrated Parallel Beam WDS/EDS Analysis: Benefits and Limitations for Geoscience Research

Giancarlo Capitani, Michael Abratis, Stefano Zanchetta, Roberto Conconi, Niccolò Magnani
article en

Abstract

The performance of an integrated parallel beam wavelength-dispersive/energy-dispersive microanalysis system mounted on a field emission scanning electron microscope has been compared with that of a conventional electron microprobe, using the same reference samples and as close as possible operational conditions. Tests on standards of known composition reveal internally consistent results for the former, and no obvious advantage of wavelength-dispersive over energy-dispersive spectroscopy has emerged for major elements without peak overlaps. Rather, because an energy-dispersive spectrometer can acquire all energies of the X-ray spectrum concurrently, conveying a lower electron dose per atom, it has the advantage of being faster and milder to minerals. In all cases, however, element migration may be significant, especially for monovalent, large ionic radius elements such as K and Na.

Microscopy and MicroanalysisVol. 31(5)
Bruker (Germany) (DE), Unité Matériaux et Transformations (FR), University of Milano-Bicocca (IT)
Openalex Percentile: Top 70%
Electron and X-Ray Spectroscopy Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Integrated Parallel Beam WDS/EDS Analysis: Benefits and Limitations for Geoscience Research — Giancarlo Capitani, Michael Abratis, et al. · Microscopy and Microanalysis (2025) | TGRS Research Map | TGRS