Influence of beam hardening correction (BHC) on Hounsfield unit accuracy: Limitations of using Teflon as a bone-like material

The application of BHC did not significantly alter HU values in low-electron-density materials such as Lung (inhale), Liver, and Adipose, with mean variations of -3.55 HU for Lung and +2.02 HU for Liver in the CIRS phantom. However, high-electron-density materials, including Teflon®, Bone 200, and Bone 800, exhibited significant HU variations (p ≤ 0.002), with HU differences ranging from -35.0 HU to -80.3 HU. The largest variation was observed in Teflon® (-80.3 HU), whereas Bone 200 and Bone 800 showed changes of -18.3 HU and -11.3 HU, respectively. Additionally, materials located in the central region of the CIRS phantom demonstrated greater HU variation, with the Liver showing a HU difference of 14.9 HU. BHC significantly influences HU values in high-electron-density materials, with the extent of this effect depending on the physical and chemical properties of the materials. Teflon® exhibited the largest HU variation, reinforcing its inadequacy for mimicking dense bone, while Delrin® displayed smaller but still notable variations. Materials from the CIRS 062M phantom showed variations within clinically acceptable limits, highlighting the importance of selecting more representative materials for accurate tissue simulation. These findings emphasize the necessity of using phantoms with materials such as those in CIRS 062M to ensure consistent and reliable HU measurements in CT imaging.

Authors

Institutions

Publication Details

Journal
Zeitschrift für Medizinische Physik
Published
2025-10-01
DOI
https://doi.org/10.1016/j.zemedi.2025.09.003
Primary Topic
Orthopaedic implants and arthroplasty
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Influence of beam hardening correction (BHC) on Hounsfield unit accuracy: Limitations of using Teflon as a bone-like material

Rafael Figueiredo Pohlmann Simões, Fernando Mecca Augusto, Thiago Lima, Paulo Vitor Pellozo
Zeitschrift für Medizinische Physik
Orthopaedic implants and arthroplasty
article

Influence of beam hardening correction (BHC) on Hounsfield unit accuracy: Limitations of using Teflon as a bone-like material

Rafael Figueiredo Pohlmann Simões, Fernando Mecca Augusto, Thiago Lima, Paulo Vitor Pellozo
article en

Abstract

The application of BHC did not significantly alter HU values in low-electron-density materials such as Lung (inhale), Liver, and Adipose, with mean variations of -3.55 HU for Lung and +2.02 HU for Liver in the CIRS phantom. However, high-electron-density materials, including Teflon®, Bone 200, and Bone 800, exhibited significant HU variations (p ≤ 0.002), with HU differences ranging from -35.0 HU to -80.3 HU. The largest variation was observed in Teflon® (-80.3 HU), whereas Bone 200 and Bone 800 showed changes of -18.3 HU and -11.3 HU, respectively. Additionally, materials located in the central region of the CIRS phantom demonstrated greater HU variation, with the Liver showing a HU difference of 14.9 HU. BHC significantly influences HU values in high-electron-density materials, with the extent of this effect depending on the physical and chemical properties of the materials. Teflon® exhibited the largest HU variation, reinforcing its inadequacy for mimicking dense bone, while Delrin® displayed smaller but still notable variations. Materials from the CIRS 062M phantom showed variations within clinically acceptable limits, highlighting the importance of selecting more representative materials for accurate tissue simulation. These findings emphasize the necessity of using phantoms with materials such as those in CIRS 062M to ensure consistent and reliable HU measurements in CT imaging.

Zeitschrift für Medizinische Physik
University of Lucerne (CH), Instituto Nacional de Câncer - INCA (BR)
Openalex Percentile: Top 51%
Orthopaedic implants and arthroplasty
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.