Effects of using auto flow factor function on hemodialysis efficacy in regular hemodialysis patients
The present study demonstrates that the use of the Auto Flow Factor Function does not result in a statistically significant difference in dialysis efficacy compared to a fixed dialysate flow rate of 500 mL/min. Given the potential for dialysate and energy savings, the implementation of Auto Flow may represent a step toward environmentally sustainable dialysis without compromising patient care.
Authors
- Mohammed Fathy Ragab
- Said Sayed Ahmed Khamis (ORCID: https://orcid.org/0000-0002-4756-0799)
- Ahmed Tawfeek (ORCID: https://orcid.org/0000-0003-2409-0517)
- Heba Kamal Abd elkhalk
Institutions
- Menoufia University (EG)
Publication Details
- Journal
- BMC Nephrology
- Published
- 2025-10-27
- DOI
- https://doi.org/10.1186/s12882-025-04508-2
- Primary Topic
- Dialysis and Renal Disease Management
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Science and Technology Development Fund