Potential of a Novel Large Left Atrial Appendage Closure Device: Short‐Term Outcomes of the WATCHMAN FLX Pro 40‐mm Versus 35‐mm

Key Points The novel WATCHMAN FLX Pro 40‐mm device enables closure of larger left atrial appendages with procedural safety comparable to that of the 35‐mm device. During the transition from the WATCHMAN FLX to the WATCHMAN FLX Pro, the overall use of large devices (≥ 35‐mm) remained similar between groups (WATCHMAN FLX: 35‐mm, 35%; WATCHMAN FLX Pro: 35‐mm, 25%; 40‐mm, 10%). The 40‐mm device was more frequently implanted in male patients and in those with non‐paroxysmal atrial fibrillation. However, the largest 40‐mm device was associated with a higher incidence of significant peri‐device leakage both at the time of the procedure and at the 3‐month follow‐up, suggesting careful patient selection and procedural planning are warranted.

Authors

Institutions

Publication Details

Journal
Catheterization and Cardiovascular Interventions
Published
2025-10-26
DOI
https://doi.org/10.1002/ccd.70282
Citations
1
Primary Topic
Atrial Fibrillation Management and Outcomes
Type
article
Field-Weighted Citation Impact
1.39
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Potential of a Novel Large Left Atrial Appendage Closure Device: Short‐Term Outcomes of the WATCHMAN FLX Pro 40‐mm Versus 35‐mm

Haruka Harata, Kazunori Mushiake, Ryuki Chatani, Hiroshi Tasaka et al.
1 citations
Catheterization and Cardiovascular Interventions
Atrial Fibrillation Management and Outcomes
1.39
article

Potential of a Novel Large Left Atrial Appendage Closure Device: Short‐Term Outcomes of the WATCHMAN FLX Pro 40‐mm Versus 35‐mm

Haruka Harata, Kazunori Mushiake, Ryuki Chatani, Hiroshi Tasaka, Takeshi Maruo, Naoki Nishiura, Shunsuke Kubo, Kazushige Kadota, Sachiyo Ono, Mikitaka Fujita
article en
1 citations

Abstract

Key Points The novel WATCHMAN FLX Pro 40‐mm device enables closure of larger left atrial appendages with procedural safety comparable to that of the 35‐mm device. During the transition from the WATCHMAN FLX to the WATCHMAN FLX Pro, the overall use of large devices (≥ 35‐mm) remained similar between groups (WATCHMAN FLX: 35‐mm, 35%; WATCHMAN FLX Pro: 35‐mm, 25%; 40‐mm, 10%). The 40‐mm device was more frequently implanted in male patients and in those with non‐paroxysmal atrial fibrillation. However, the largest 40‐mm device was associated with a higher incidence of significant peri‐device leakage both at the time of the procedure and at the 3‐month follow‐up, suggesting careful patient selection and procedural planning are warranted.

Catheterization and Cardiovascular InterventionsVol. 106(7)
Kurashiki Central Hospital (JP)
Openalex Percentile: Top 14%
Atrial Fibrillation Management and Outcomes
1.39
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.