Ventricular Tachycardia Due to Blanking of a Premature Ventricular Contraction
A 78-year-old man was evaluated for a non-sustained ventricular tachycardia episode detected in a dual-chamber pacemaker (Boston Scientific Essentio MRI EL L131) implanted for sick sinus syndrome. Blanking of a premature ventricular beat caused delivery of ventricular pacing in the vulnerable period triggering non-sustained ventricular tachycardia. Review of literature found more than half of reported cases involved Boston devices. The commonalities of such phenomenon are summarized and reprogramming strategies presented.
Authors
- Shing Ching (ORCID: https://orcid.org/0000-0001-9805-3191)
- Alexson Tsz‐Ki Lau
Institutions
- United Christian Hospital (CN)
Publication Details
- Journal
- Pacing and Clinical Electrophysiology
- Published
- 2025-10-25
- DOI
- https://doi.org/10.1111/pace.70075
- Primary Topic
- Cardiac pacing and defibrillation studies
- Type
- article
- Field-Weighted Citation Impact
- 0.00