S485 Qualitative Comparison of Air-Charged and Solid-State HRAM Catheter Normative Values in Healthy Subjects

Introduction: While high-resolution anorectal manometry (HRAM) is essential for evaluating anorectal function, absolute pressure values can vary substantially depending on the catheter type. Air-charged and solid-state catheters differ in construction, transduction mechanism, and calibration method—each influencing recorded values. This variability necessitates catheter-specific interpretation. As a secondary objective of the ILLUMINATE study, we compared normative data from a novel 10-Channel Air-Charged Disposable HRAM Catheter (Solar catheter) to published values from solid-state systems to assess qualitative agreement and highlight device-specific considerations. Methods: The ILLUMINATE study was a prospective, multi-center, open-label trial wherein all subjects underwent standardized HRAM using a novel 10-channel air-charged disposable HRAM catheter per International Anorectal Physiology Working Group protocol and London Classification, and a subset also underwent testing with the solid-state (UniTip) catheter (n = 30). Parameters collected included: anal resting pressure, squeeze pressure, cough pressure, and functional anal canal length. Values were descriptively compared to published normative datasets from solid-state catheter systems. Results: There were 30 subjects (3 men, 37 women; mean age = 39 ± 9 years) who underwent both novel 10-channel air-charged disposable catheter and solid-state catheter ARM studies. Anal resting pressures recorded with the air-charged catheter were approximately equivalent to solid-state reference values. Functional anal canal length measurements were consistent across technologies as well. Conclusion: Minor differences were observed between air-charged and solid-state HRAM catheters, but in general physiologic trends were preserved across systems. These findings underscore the importance of using catheter-specific normative values when interpreting HRAM results and reinforce that direct numeric comparison between devices may not be valid. The novel 10-Channel Air-Charged Disposable HRAM catheter demonstrated consistent performance and clinical utility within its calibrated reference range.

Authors

Publication Details

Journal
The American Journal of Gastroenterology
Published
2025-10-01
DOI
https://doi.org/10.14309/01.ajg.0001129400.07879.e4
Primary Topic
Diagnosis and Treatment of Venous Diseases
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

S485 Qualitative Comparison of Air-Charged and Solid-State HRAM Catheter Normative Values in Healthy Subjects

Kamran Ayub, Allan P. Weston, Delaram Asadi, Danielle Long et al.
The American Journal of Gastroenterology
Diagnosis and Treatment of Venous Diseases
article

S485 Qualitative Comparison of Air-Charged and Solid-State HRAM Catheter Normative Values in Healthy Subjects

Kamran Ayub, Allan P. Weston, Delaram Asadi, Danielle Long, Satish S. Rao, Karlo Fidel
article en

Abstract

Introduction: While high-resolution anorectal manometry (HRAM) is essential for evaluating anorectal function, absolute pressure values can vary substantially depending on the catheter type. Air-charged and solid-state catheters differ in construction, transduction mechanism, and calibration method—each influencing recorded values. This variability necessitates catheter-specific interpretation. As a secondary objective of the ILLUMINATE study, we compared normative data from a novel 10-Channel Air-Charged Disposable HRAM Catheter (Solar catheter) to published values from solid-state systems to assess qualitative agreement and highlight device-specific considerations. Methods: The ILLUMINATE study was a prospective, multi-center, open-label trial wherein all subjects underwent standardized HRAM using a novel 10-channel air-charged disposable HRAM catheter per International Anorectal Physiology Working Group protocol and London Classification, and a subset also underwent testing with the solid-state (UniTip) catheter (n = 30). Parameters collected included: anal resting pressure, squeeze pressure, cough pressure, and functional anal canal length. Values were descriptively compared to published normative datasets from solid-state catheter systems. Results: There were 30 subjects (3 men, 37 women; mean age = 39 ± 9 years) who underwent both novel 10-channel air-charged disposable catheter and solid-state catheter ARM studies. Anal resting pressures recorded with the air-charged catheter were approximately equivalent to solid-state reference values. Functional anal canal length measurements were consistent across technologies as well. Conclusion: Minor differences were observed between air-charged and solid-state HRAM catheters, but in general physiologic trends were preserved across systems. These findings underscore the importance of using catheter-specific normative values when interpreting HRAM results and reinforce that direct numeric comparison between devices may not be valid. The novel 10-Channel Air-Charged Disposable HRAM catheter demonstrated consistent performance and clinical utility within its calibrated reference range.

The American Journal of GastroenterologyVol. 120(10S2)
Openalex Percentile: Top 52%
Diagnosis and Treatment of Venous Diseases
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.