Evaluation of different techniques of measurement of the axial length in silicone oil-filled eyes

Purpose To evaluate different techniques of axial length measurement in silicone oil-filled eyes and compare the results of each with the original axial length measured before pars plana vitrectomy and silicone oil injection for accurate measurement of intraocular lens calculation in these eyes. Patients and methods This retrospective observational cohort case series study included 30 eyes of 30 patients who underwent pars plana vitrectomy and silicone oil injection. The axial length of the silicone-filled eye was measured by A-scan ultrasonography using two different ultrasound velocities: the normal ultrasound wave velocity, an adjusted velocity and the correcting factor. The axial length was also measured by optical biometry and compared to the original actual axial length. Results The mean preoperative axial length by A-scan biometry was 23.33±1.37 mm. The postoperative axial length by standard velocity multiplied by the correcting factor of 0.71 was 23.21±1.25 mm. The postoperative axial length, using a slower velocity of 1000 m/sec, was 21.06±1.34 mm. The axial length measured by optical biometry using a standard program postoperatively was 24.27±1.2 mm, which had the highest intraclass correlation coefficient of 0.948 and Cronbach’s alpha 0.973. The postoperative axial length measured by optical biometry of the silicone-specialized program was 23.1±1.43 mm. The difference between measurements of postoperative axial length, compared to the preoperative value, was statistically significant ( P <0.001). Conclusion The axial length of the silicone oil-filled eye by the standard program of optical biometry was the most accurate method compared to the preoperative axial length measured by A-scan ultrasonography. It can also be accurately measured by the silicone-specialized program of optical biometry. A-scan ultrasonography of the silicone oil-filled eye using the standard ultrasound wave velocity (1548 m/sec) and the correcting factor gives results almost as accurate as the optical biometry using the silicone-specialized program.

Authors

Institutions

Publication Details

Journal
Journal of the Egyptian Ophthalmological Society
Published
2025-10-01
DOI
https://doi.org/10.4103/ejos.ejos_40_25
Primary Topic
Ophthalmology and Visual Impairment Studies
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Evaluation of different techniques of measurement of the axial length in silicone oil-filled eyes

Ahmed A. Mohalhal, K.C. Joseph, Ghada A. Nassar, Ehsan El-Mehelmy
Journal of the Egyptian Ophthalmological Society
Ophthalmology and Visual Impairment Studies
article

Evaluation of different techniques of measurement of the axial length in silicone oil-filled eyes

Ahmed A. Mohalhal, K.C. Joseph, Ghada A. Nassar, Ehsan El-Mehelmy
article en

Abstract

Purpose To evaluate different techniques of axial length measurement in silicone oil-filled eyes and compare the results of each with the original axial length measured before pars plana vitrectomy and silicone oil injection for accurate measurement of intraocular lens calculation in these eyes. Patients and methods This retrospective observational cohort case series study included 30 eyes of 30 patients who underwent pars plana vitrectomy and silicone oil injection. The axial length of the silicone-filled eye was measured by A-scan ultrasonography using two different ultrasound velocities: the normal ultrasound wave velocity, an adjusted velocity and the correcting factor. The axial length was also measured by optical biometry and compared to the original actual axial length. Results The mean preoperative axial length by A-scan biometry was 23.33±1.37 mm. The postoperative axial length by standard velocity multiplied by the correcting factor of 0.71 was 23.21±1.25 mm. The postoperative axial length, using a slower velocity of 1000 m/sec, was 21.06±1.34 mm. The axial length measured by optical biometry using a standard program postoperatively was 24.27±1.2 mm, which had the highest intraclass correlation coefficient of 0.948 and Cronbach’s alpha 0.973. The postoperative axial length measured by optical biometry of the silicone-specialized program was 23.1±1.43 mm. The difference between measurements of postoperative axial length, compared to the preoperative value, was statistically significant ( P <0.001). Conclusion The axial length of the silicone oil-filled eye by the standard program of optical biometry was the most accurate method compared to the preoperative axial length measured by A-scan ultrasonography. It can also be accurately measured by the silicone-specialized program of optical biometry. A-scan ultrasonography of the silicone oil-filled eye using the standard ultrasound wave velocity (1548 m/sec) and the correcting factor gives results almost as accurate as the optical biometry using the silicone-specialized program.

Journal of the Egyptian Ophthalmological SocietyVol. 118(4)
Cairo University (EG), Heliopolis University (EG)
Openalex Percentile: Top 59%
Ophthalmology and Visual Impairment Studies
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.