Valve-In-Valve TAVI in a Patient With Dextrocardia, Situs Inversus, and Challenging Commissural Alignment

Valve-in-valve transcatheter aortic valve implantation (ViV-TAVI) for dextrocardia is challenging. An 81-year-old man with dextrocardia and situs inversus developed prosthetic aortic stenosis. Loading Evolut FX+ C-tab 180° upside down in a delivery catheter facilitates commissural alignment during ViV-TAVI in dextrocardia. To accommodate anatomical variations, the flush-port position can be adjusted. In this case, the flush port was positioned 5-o'clock (60° from the usual 3-o'clock position) to align the TAVI valve with former prosthesis, as the former prosthetic valve commissure was oriented 60° from native commissures to preserve coronary access for abnormal left coronary artery originating from native commissures. This case illustrates strategies for commissural alignment during ViV-TAVI in dextrocardia.

Authors

Institutions

Publication Details

Journal
JACC Case Reports
Published
2025-10-25
DOI
https://doi.org/10.1016/j.jaccas.2025.105833
Primary Topic
Cardiac Valve Diseases and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Valve-In-Valve TAVI in a Patient With Dextrocardia, Situs Inversus, and Challenging Commissural Alignment

Daniel Tai‐Leung Chan, Stephen Cho Wing Sze, Chor Cheung Tam, Simon Cheung-Chi Lam et al.
JACC Case Reports
Cardiac Valve Diseases and Treatments
article

Valve-In-Valve TAVI in a Patient With Dextrocardia, Situs Inversus, and Challenging Commissural Alignment

Daniel Tai‐Leung Chan, Stephen Cho Wing Sze, Chor Cheung Tam, Simon Cheung-Chi Lam, Chun‐Ka Wong, Ka‐Chun Un, K. Chan, Ching-Wei Lee, Gilbert Tang, Ho-On Alston Conrad Chiu
article en

Abstract

Valve-in-valve transcatheter aortic valve implantation (ViV-TAVI) for dextrocardia is challenging. An 81-year-old man with dextrocardia and situs inversus developed prosthetic aortic stenosis. Loading Evolut FX+ C-tab 180° upside down in a delivery catheter facilitates commissural alignment during ViV-TAVI in dextrocardia. To accommodate anatomical variations, the flush-port position can be adjusted. In this case, the flush port was positioned 5-o'clock (60° from the usual 3-o'clock position) to align the TAVI valve with former prosthesis, as the former prosthetic valve commissure was oriented 60° from native commissures to preserve coronary access for abnormal left coronary artery originating from native commissures. This case illustrates strategies for commissural alignment during ViV-TAVI in dextrocardia.

JACC Case ReportsVol. 30(40)
Twitter (United States) (US), Queen Mary Hospital (CN), Grantham Hospital (CN), Taipei Veterans General Hospital (TW), University of Hong Kong (HK)
Openalex Percentile: Top 54%
Cardiac Valve Diseases and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.