Determining the effective energy and scattered radiation distributions in an interventional cardiology department

This study aims to determine effective energy and scattered radiation distributions in an interventional cardiology department. Additionally, the protective effectiveness of ceiling-suspended shielding was investigated. Optically stimulated luminescence dosemeters were attached to mica rods to determine effective energy and scattered radiation at numerous points in the catheterization laboratory. The effective energy of the scattered beam ranged from 40 to 50 keV, with the beam scattered to the eye position reaching an effective energy of 45 keV. Radiation was distributed nonuniformly at different locations and heights. Ceiling-suspended shielding reduced the dose to the head and neck region but did not significantly affect other regions. The combination of a suitable standing position and ceiling-suspended shielding can notably diminish the staff's eye dose by >93%.

Authors

Institutions

Publication Details

Journal
Radiation Protection Dosimetry
Published
2025-10-03
DOI
https://doi.org/10.1093/rpd/ncaf125
Primary Topic
Radiation Dose and Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Determining the effective energy and scattered radiation distributions in an interventional cardiology department

Trung-Dung Hoang, Lena Dang, N. Nguyen, Minh Tan Nguyen et al.
Radiation Protection Dosimetry
Radiation Dose and Imaging
article

Determining the effective energy and scattered radiation distributions in an interventional cardiology department

Trung-Dung Hoang, Lena Dang, N. Nguyen, Minh Tan Nguyen, Hai H. Bui, Liem Thanh Le, Tuyen Nhu Pham
article en

Abstract

This study aims to determine effective energy and scattered radiation distributions in an interventional cardiology department. Additionally, the protective effectiveness of ceiling-suspended shielding was investigated. Optically stimulated luminescence dosemeters were attached to mica rods to determine effective energy and scattered radiation at numerous points in the catheterization laboratory. The effective energy of the scattered beam ranged from 40 to 50 keV, with the beam scattered to the eye position reaching an effective energy of 45 keV. Radiation was distributed nonuniformly at different locations and heights. Ceiling-suspended shielding reduced the dose to the head and neck region but did not significantly affect other regions. The combination of a suitable standing position and ceiling-suspended shielding can notably diminish the staff's eye dose by >93%.

Radiation Protection DosimetryVol. 201(19)
Trường ĐH Nguyễn Tất Thành (VN), Central Military Hospital (NL), CM Hospital (KR)
Openalex Percentile: Top 55%
Radiation Dose and Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Determining the effective energy and scattered radiation distributions in an interventional cardiology department — Trung-Dung Hoang, Lena Dang, et al. · Radiation Protection Dosimetry (2025) | TGRS Research Map | TGRS