Automated alignment of XFEL nanofocusing mirrors via wavefront optimization
An automated alignment procedure, based on wavefront measurement with a single-grating interferometer, has been developed for precise tuning of Kirkpatrick-Baez nanofocusing mirrors for X-ray free-electron lasers (XFELs). This approach optimizes focus size and maximizes peak intensity while minimizing aberrations. Wavefront errors are quantitatively correlated with alignment deviations - incidence angle, perpendicularity and astigmatism - via Legendre polynomial analysis. These errors are subsequently corrected through a straightforward optimization process. Implemented at the SPring-8 Angstrom Compact Free-Electron Laser (SACLA), the system consistently achieves a reproducible XFEL focus below 150 nm × 200 nm within 10 min. Routine operation at SACLA demonstrates the reliability and efficacy of this method, enabling rapid restoration of optimal nanofocusing conditions.
Authors
- Taito Osaka (ORCID: https://orcid.org/0000-0003-0545-9737)
- Ichiro Inoue (ORCID: https://orcid.org/0009-0008-0423-1347)
- Jumpei Yamada (ORCID: https://orcid.org/0000-0002-5476-4327)
- Gota Yamaguchi (ORCID: https://orcid.org/0000-0003-4305-807X)
- Makina Yabashi (ORCID: https://orcid.org/0000-0002-2472-1684)
- Yuichi Inubushi
Institutions
- SPring-8 (JP)
- Japan Synchrotron Radiation Research Institute (JP)
- The University of Osaka (JP)
Publication Details
- Journal
- Journal of Synchrotron Radiation
- Published
- 2025-10-09
- DOI
- https://doi.org/10.1107/s1600577525008434
- Primary Topic
- Advanced X-ray Imaging Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00
Funders
- Japan Society for the Promotion of Science
- Japan Science and Technology Agency
- RIKEN
- Precursory Research for Embryonic Science and Technology