Intraprosthetic Dislocation Following Reduction of Dual-mobilityTotal Hip Arthroplasty
Intraprosthetic dislocation is a complication specific to dual-mobility hip prosthetics, characterized by displacement of the polyurethane liner unique to this type of device. This liner was designed to offer some benefits over other types of hip prosthetics, including improved biomechanics and lower risk of dislocation. The liner can become dislodged during a hip dislocation and remain displaced despite successful reduction of the metallic prosthetic components. Due to its radiolucency, diagnosis can be challenging on plain radiographs, often requiring advanced imaging. This case highlights the importance of recognizing this potential complication in patients with dual-mobility prostheses.
Authors
- Matthias Barden (ORCID: https://orcid.org/0000-0002-1731-4933)
- Marissa Benbassat
- Emilio Benbassat
Institutions
- Eisenhower Medical Center (US)
Publication Details
- Journal
- Clinical Practice and Cases in Emergency Medicine
- Published
- 2025-10-24
- DOI
- https://doi.org/10.5811/cpcem.47211
- Primary Topic
- Orthopaedic implants and arthroplasty
- Type
- article
- Field-Weighted Citation Impact
- 0.00