The Hyphenation of High-Performance Liquid Chromatography with X-ray Fluorescence for Universal, Flow-Through, Elemental Analysis of Organobromines

Current detectors used in liquid chromatography are still lacking in what is desired of an ideal detector. One of those characteristics is the universal response. In this work, the novel use of X-ray fluorescence as a complementary, flow-through, universal detector was developed in hyphenation with reversed-phase chromatography. Monte Carlo simulations were performed to aid in the creation of the HPLC-XRF setup and study the influence of parameters involved. A proof of principle was established for the analysis of organobromines of environmental concern, and calibration curves were created, which show LODs below 0.3 mM for all tested compounds and good linearity between 0.2 and 20 mM (<i>R</i><sup>2</sup> > 0.99). The relative standard deviation of the peak areas varied between 10.47 and 1.06% with an average of 3.67%. The universal character of XRF was shown via flow injection analysis, allowing the use of element-based calibration curves instead of compound-specific ones. This conclusion can be further extended to other elements detectable by XRF. Sensitivity is currently a limitation but can be further improved through multidetector setups, improved acquisition hardware and software, and techniques such as stop-flow chromatography and preconcentration. The isocratic and gradient separation of brominated species was also performed to illustrate the potential of HPLC-XRF in separation studies. While simulations predicted a minor influence of mobile phase composition on the response, this effect proved to be negligible under the experimental conditions.

Authors

Institutions

Publication Details

Journal
Analytical Chemistry
Published
2025-10-24
DOI
https://doi.org/10.1021/acs.analchem.5c03394
Primary Topic
Analytical chemistry methods development
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

The Hyphenation of High-Performance Liquid Chromatography with X-ray Fluorescence for Universal, Flow-Through, Elemental Analysis of Organobromines

Gaëlle Spileers, Pieter Tack, Frédéric Lynen, László Vincze
Analytical Chemistry
Analytical chemistry methods development
article

The Hyphenation of High-Performance Liquid Chromatography with X-ray Fluorescence for Universal, Flow-Through, Elemental Analysis of Organobromines

Gaëlle Spileers, Pieter Tack, Frédéric Lynen, László Vincze
article en

Abstract

Current detectors used in liquid chromatography are still lacking in what is desired of an ideal detector. One of those characteristics is the universal response. In this work, the novel use of X-ray fluorescence as a complementary, flow-through, universal detector was developed in hyphenation with reversed-phase chromatography. Monte Carlo simulations were performed to aid in the creation of the HPLC-XRF setup and study the influence of parameters involved. A proof of principle was established for the analysis of organobromines of environmental concern, and calibration curves were created, which show LODs below 0.3 mM for all tested compounds and good linearity between 0.2 and 20 mM (<i>R</i><sup>2</sup> > 0.99). The relative standard deviation of the peak areas varied between 10.47 and 1.06% with an average of 3.67%. The universal character of XRF was shown via flow injection analysis, allowing the use of element-based calibration curves instead of compound-specific ones. This conclusion can be further extended to other elements detectable by XRF. Sensitivity is currently a limitation but can be further improved through multidetector setups, improved acquisition hardware and software, and techniques such as stop-flow chromatography and preconcentration. The isocratic and gradient separation of brominated species was also performed to illustrate the potential of HPLC-XRF in separation studies. While simulations predicted a minor influence of mobile phase composition on the response, this effect proved to be negligible under the experimental conditions.

Analytical ChemistryVol. 97(43)
HOGENT University of Applied Sciences and Arts (BE), Ghent University Hospital (BE), Ghent University (BE)
Openalex Percentile: Top 65%
Analytical chemistry methods development
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.